Current through Register No. 40, October 3, 2024
(a)
When the x-ray tube is operated at its maximum rated tube current for the
maximum kV, the leakage air kerma rate shall not exceed the value specified at
the distance specified as follows:
(1) For 5
to 50 kV systems, the leakage air kerma rate measured at any position 5
centimeters from the tube housing assembly shall not exceed 1 mGy (100 mrad) in
any one-hour period;
(2) For >50
and <500 kV systems:
a. The leakage air
kerma rate measured at a distance of 1 meter from the target in any direction
shall not exceed 1 cGy (1 rad) in any one-hour period;
b. This air kerma rate measurement may be
averaged over areas no larger than 100 square centimeters; and
c. The air kerma rate at a distance of 5
centimeters from the surface of the tube housing assembly shall not exceed 30
cGy (30 rad) per hour;
(3) For each therapeutic radiation machine,
the registrant shall determine, or obtain from the manufacturer, the leakage
radiation existing at the positions specified in
He-P
4047.06(a) (1) and (a) (2), for the
specified operating conditions; and
(4) Records on leakage radiation measurements
shall be maintained after installation for inspection by DHHS/RHS.
(b) Permanent diaphragms or cones
used for limiting the useful beam shall provide at least the same degree of
attenuation as required for the tube housing assembly.
(c) All adjustable or removable beam limiting
devices, diaphragms, cones or blocks shall meet the following requirements:
(1) None shall transmit more than 5 percent
of the useful beam for the most penetrating beam used; and
(2) When adjustable or removable beam
limiting devices are used, the position and shape of the radiation field shall
be indicated by a light beam.
(d) The filter system shall be so designed
that:
(1) Filters can not be accidentally
displaced at any possible tube orientation;
(2) An interlock system prevents irradiation
if the proper filter is not in place;
(3) The air kerma rate escaping from the
filter slot shall not exceed 1 cGy (1 rad) per hour at one meter under any
operating conditions; and
(4) Each
filter shall be marked as to its material of construction and its
thickness.
(e) The x-ray
tube shall be so mounted that it can not accidentally turn or slide with
respect to the housing aperture.
(f) The tube housing assembly shall be
capable of being immobilized for stationary portal treatments.
(g) The tube housing assembly shall be marked
so that:
(1) It is possible to determine the
location of the source to within 5 millimeters; and
(2) Such marking shall be readily accessible
for use during calibration procedures.
(h) Contact therapy tube housing assemblies
shall have a removable shield of material, equivalent in attenuation to 0.5
millimeters of lead at 100 kV, which can be positioned over the entire useful
beam exit port during periods when the beam is not in use.
(i) A timer control device shall be provided
to terminate the irradiation after a pre-set time interval and shall:
(1) Have a display;
(2) Be provided at the treatment
control;
(3) Have a pre-set time
selector;
(4) Have elapsed time or
time remaining indicator;
(5) Be a
cumulative timer which activates with an indication of "BEAM-ON" and retains
its reading after irradiation is interrupted or terminated;
(6) Be able to reset the elapsed time
indicator after irradiation is terminated and before irradiation can be
re-initiated;
(7) Terminate
irradiation when a pre-selected time has elapsed, if any dose monitoring system
present has not previously terminated irradiation;
(8) Permit accurate pre-setting and
determination of exposure times as short as 1 second;
(9) Not permit an exposure if set at
zero;
(10) Not activate until the
shutter is opened when irradiation is controlled by a shutter mechanism unless
calibration includes a timer error correction to compensate for mechanical lag;
and
(11) Be accurate to within 1
percent of the selected value or 1 second, whichever is greater.
(j) The control panel, in addition
to the displays required by other provisions in
He-P
4047.06, shall have:
(1) An indication of whether electrical power
is available at the control panel and if activation of the x-ray tube is
possible;
(2) An indication of
whether x-rays are being produced;
(3) Means for indicating x-ray tube potential
and current;
(4) Means for
terminating exposure at any time;
(5) A locking device which will prevent
unauthorized use of the therapeutic radiation machine; and
(6) A positive display of specific filter(s)
in the beam.
(k) A
control panel which may energize more than one x-ray tube shall:
(1) Activate only one x-ray tube at any
time;
(2) Have an indication at the
control panel identifying which x-ray tube is activated; and
(3) Have an indication at the tube housing
assembly when that tube is energized.
(l) There shall be a means of determining the
central axis target-to-skin distance to within 1 centimeter and of reproducing
this measurement to within 2 millimeters thereafter.
(m) Shutters shall be required as follows:
(1) Unless it is possible to bring the x-ray
output to the prescribed exposure parameters within 5 seconds after the x-ray
"ON" switch is energized, the beam shall be attenuated by a shutter having a
lead equivalency not less than that of the tube housing assembly;
(2) After the unit is at operating
parameters, the shutter shall be controlled by the operator from the control
panel; and
(3) An indication of
shutter position shall appear at the control panel.
(n) Each therapeutic radiation machine
equipped with a beryllium or other low-filtration window shall:
(1) Be clearly labeled as such on the tube
housing assembly; and
(2) Be
provided with a permanent warning device on the control panel that is activated
when there is no additional filtration present, in order to indicate that the
dose rate is very high.
(o) In addition to the shielding requirements
of He-P 4047, the treatment room shall meet the following facility design
requirements for therapeutic radiation machines capable of operating in the
range 50 kV to 500 kV:
(1) Provision shall be
made for continuous two-way aural communication between the patient and the
operator at the control panel; and
(2) A viewing system shall:
a. Be provided to permit continuous
observation of the patient during irradiation;
b. Enable the operator to observe the patient
from the control panel; and
c. Be
operational for use of therapeutic radiation machine for patient
irradiation.
(p) Treatment rooms which contain a
therapeutic radiation machine capable of operating above 150 kV shall meet the
following additional requirements:
(1) All
protective barriers shall be fixed except for entrance doors or beam
interceptors;
(2) The control panel
shall be located outside the treatment room or in a totally enclosed booth,
which has a shielded ceiling, inside the room;
(3) Interlocks shall be provided such that
all entrance doors, including doors to any interior booths, shall be closed
before treatment can be initiated or continued;
(4) If the radiation beam is interrupted by
any door opening, it shall not be possible to restore the machine to operation
without closing the door and re-initiating irradiation by manual action at the
control panel; and
(5) When any
door referred to in He-P 4047.06(p)
is opened while the x-ray tube is activated,
the air kerma rate at a distance of 1 meter from the source shall be reduced to
less than 1 mGy (100 mrad) per hour.
(q) Full calibration of a therapeutic
radiation machine shall be performed by, or under the direct supervision of, a
radiation therapy physicist and shall meet the following requirements:
(1) Full calibration shall be performed as
follows:
a. Before the first medical use
following installation or reinstallation of the therapeutic radiation
machine;
b. At intervals not
exceeding 1 year; and
c. Before
medical use under the following conditions:
1. Whenever quality assurance check
measurements indicate that the radiation output differs by more than 5 percent
from the value obtained at the last full calibration and the differences cannot
be reconciled; and
2. Following any
component replacement, major repair, or modification of components that could
significantly affect the characteristics of the radiation beam;
(2) Notwithstanding the
requirements of He-P 4047.06(q) (1)
c.1. above:
a. Full calibration of therapeutic radiation
machines with multi-energy capabilities is required only for those modes and/or
energies that are not within their acceptable range;
b. If the repair, replacement, or
modification does not affect all energies, full calibration shall be performed
on the affected energy that is in most frequent clinical use at the facility;
and
c. Any remaining energies may
be validated with quality assurance check procedures against the criteria in
He-P
4047.06(q) (1) c.1;
(3) For machines with energies
greater than or equal to 10 KeV and less than or equal to 40 KeV full
calibration shall include all measurements recommended for annual calibration
by National Council on Radiation Protection (NCRP) Report 69, "Dosimetry of
X-ray and Gamma Ray Beams for Radiation Therapy in the Energy Range 10 KeV to
50 MeV" (1981), which is incorporated by reference and included in Appendix A.
For machines with energies greater than or equal to 40 kV and less than equal
to 300 kV, full calibration shall include all measurements recommended for
annual calibration by "High Dose-Rate Brachytherapy Treatment Delivery" (1998)
, AAPM Task Group 59, and "AAPM Protocol for 40-300 kV X-Ray Beam Dosimetry in
Radiotherapy and Radiobiology" (2001) by AAPM Task Group 61, both of which are
incorporated by reference and included in Appendix A;
(4) The registrant shall maintain a record of
each calibration for the duration of the registration; and
(5) The calibration record shall include:
a. The date of the calibration;
b. The manufacturer's name, model number, and
serial number for both the therapeutic machine and the x-ray tube;
c. The model numbers and serial numbers of
the instruments used to calibrate the therapeutic radiation machine;
and
d. The signature of the
radiation therapy physicist responsible for performing the
calibration.
(r) Periodic quality assurance checks shall
be performed on therapeutic radiation machines subject to
He-P
4047.06, which are capable of operation at greater
than or equal to 50 kV, as follows:
(1) The
registrant shall perform quality assurance checks specified in
He-P
4047.06(r) in accordance with
written procedures established by the radiation therapy physicist;
(2) The quality assurance check procedures
required by He-P 4047.06(r) (1)
shall specify:
a. The frequency at which tests or
measurements are to be performed;
b. That the quality assurance check be
performed during the calibration specified in
He-P
4047.06(q) ; and
c. The acceptable tolerance for each
parameter measured in the quality assurance check, when compared to the value
for that parameter determined in the calibration as specified by
He-P
4047.06(q) (1) ;
(3) The quality assurance check shall
investigate the cause for a parameter exceeding a tolerance set by the
radiation therapy physicist and correct such parameter before the system is
used for patient irradiation;
(4)
Whenever a quality assurance check indicates a significant change in the
operating characteristics of a system, as specified in the radiation therapy
physicist's quality assurance check procedures, the system shall be
recalibrated as specified by
He-P
4047.06(q) (1) ;
(5) The registrant shall use the dosimetry
system described in He-P 4047.05(k)
to make the quality assurance check required
in He-P
4047.06(r) (2) ;
(6) The registrant shall have the radiation
therapy physicist review and sign the results of each radiation output quality
assurance check within 30 days of the date that the check was
performed;
(7) The registrant shall
ensure that the safety quality assurance checks of therapeutic radiation
machines are performed at intervals not to exceed 30 days;
(8) The registrant shall ensure that no
therapeutic machine is used to administer radiation to humans unless the
quality assurance checks will have been performed within the 30-day period
immediately prior to said administration;
(9) Safety quality assurance checks shall
ensure proper operation of:
a. Electrical
interlocks at each external beam radiation therapy room entrance;
b. Proper operation of the "BEAM-ON" and
termination switches;
c. Beam
condition indicator lights on the access door(s), control console, and in the
radiation therapy room;
d. Viewing
systems; and
e. If applicable,
electrically operated treatment room doors from inside and outside the
treatment room;
(10) The
registrant shall maintain a record of each quality assurance check for 3 years;
and
(11) The quality assurance
check records shall include:
a. The date of
the quality assurance check;
b. The
manufacturer's name, model number, and serial number for the therapeutic
radiation machine;
c. The
manufacturer's name, model number, and serial number of the instrument(s) used
to measure the radiation output of the therapeutic radiation machine;
and
d. The signature of the
individual who performed the periodic quality assurance check.
(s) The following
operating procedures shall be met:
(1) The
therapeutic radiation machine shall not be used for irradiation of patients
unless the requirements of
He-P
4047.06(q) and
He-P
4047.06(r) have been met;
(2) Therapeutic radiation machines shall not
be left unattended unless secured pursuant to
He-P
4047.06(j) (5) ;
(3) When a patient must be held in position
for radiation therapy, mechanical supporting or restraining devices shall be
used;
(4) The tube housing assembly
shall not be held by an individual during operation unless the assembly is
designed to require such holding and the peak tube potential of the system does
not exceed 50 kV;
(5) If the tube
housing is held, the holder shall wear protective gloves and apron of not less
than 0.5 millimeters lead equivalency at 100 kV;
(6) A copy of the current operating and
emergency procedures shall be maintained at the therapeutic radiation machine
control console;
(7) No individual
other than the patient shall be in the treatment room during exposures from
therapeutic radiation machines operating above 150 kV; and
(8) At energies less than or equal to 150 kV,
any individual, other than the patient, in the treatment room shall be
protected by a barrier sufficient to meet the requirements of He-P
4020.
(t) Each facility
location authorized to use a therapeutic radiation machine in accordance with
He-P
4047.06 shall:
(1)
Possess appropriately calibrated portable monitoring equipment;
(2) As a minimum, include a portable
radiation measurement survey instrument capable of measuring dose rates over
the range of 10 µSv (1 mrem) per hour to 10 mSv (1000 mrem) per hour;
and
(3) Require the survey
instrument(s) to be operable and calibrated in accordance with
He-P
4047.08.
#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06,
EXPIRES: 2-2-07; ss by #8808, eff
1-24-07