New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4046 - USE OF RADIATION MACHINES: GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC AND VETERINARY X-RAY SYSTEMS
Section He-P 4046.13 - Computed Tomography X-ray Systems - Surveys, Calibrations, Spot Checks, and Operating Procedures
Current through Register No. 40, October 3, 2024
(a) All CT x-ray systems shall have a survey made by or under the direction of a qualified expert.
(b) The registrant shall obtain a written report of the survey from the qualified expert, and a copy of the report shall be made available to DHHS/RHS upon request.
(c) The calibration of the radiation output of the CT x-ray system shall be performed by or under the direction of a qualified expert who is physically present at the facility during such calibration.
(d) The calibration of a CT x-ray system shall be performed at intervals specified by a qualified expert and after any change or replacement of components which, in the opinion of the qualified expert, could cause a change in the radiation output.
(e) The calibration of the radiation output of a CT x-ray system shall be performed with a calibrated dosimetry system which is traceable to a national standard and has been calibrated within the preceding 2 years.
(f) CT dosimetry phantom(s) shall meet the following specifications and conditions of use shall be used in determining the radiation output of a CT X-ray system:
(g) The calibration shall be required for each type of head, body, or whole-body scan performed at the facility.
(h) Calibration shall meet the following requirements:
(i) Calibration procedures shall be in writing.
(j) Records of calibrations performed shall be maintained for inspection by DHHS/RHS.
(k) The spot check procedures shall be in writing and shall have been developed by a qualified expert.
(l) The spot check procedures shall incorporate the use of a CT dosimetry phantom capable of providing an indication of contrast scale, noise, nominal tomographic section thickness, resolution capability of the system for low and high contrast objects, and measuring the mean CTN for water or other reference material.
(m) All spot checks shall be included in the CT calibration and at time intervals and under system conditions specified by a qualified expert.
(n) Spot checks shall include acquisition of images obtained with the CT dosimetry phantom(s) using the same processing mode and CT conditions of operation as are used to perform calibrations.
(o) The spot check images shall be retained, until a new calibration is performed, as images stored in digital form on a backed-up storage medium compatible with the CT x-ray system.
(p) Written records of the spot checks performed shall be maintained for inspection by DHHS/RHS.
(q) The CT x-ray system shall not be operated except by a person who has been specifically trained in its operation.
(r) Information shall be available at the control panel regarding the operation and calibration of the system to include the following:
#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07