Section He-P 4046.12 - Computed Tomography X-ray Systems - Facility Design Requirements

Universal Citation: NH Admin Rules He-P 4046.12

Current through Register No. 40, October 3, 2024

(a) Provision shall be made for 2-way aural communication between the patient and the operator at the control panel.

(b) Windows, mirrors, closed-circuit television, or an equivalent shall be provided to permit continuous observation of the patient during irradiation and shall be so located that the operator can observe the patient from the control panel.

(c) When the primary viewing system is by electronic means, an alternate viewing system shall be available for use in the event of failure of the primary viewing system.

#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07

Disclaimer: These regulations may not be the most recent version. New Hampshire may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.