New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4046 - USE OF RADIATION MACHINES: GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC AND VETERINARY X-RAY SYSTEMS
Section He-P 4046.09 - Hand-held Dental Intraoral Machines

Universal Citation: NH Admin Rules He-P 4046.09

Current through Register No. 40, October 3, 2024

The following requirements shall apply to the use of hand-held dental intraoral machines:

(a) For all uses:

(1) The registrant shall follow all applicable requirements set forth in the New Hampshire Rules for the Control of Radiation (NHRCR), as well as all manufacturers's requirements;

(2) The machine shall be used only for dental diagnostic imaging;

(3) The protective shielding shall not be removed or modified;

(4) The registrant shall provide a copy of their operating, safety and security procedures to operators to prevent unauthorized or improper use;

(5) The registrant shall prepare a quality assurance plan, with an emphasis on ensuring limited retaking of radiographs;

(6) Operators of hand-held dental intraoral machines shall be specifically trained to operate such equipment, and the registrant shall provide proof of training for all operators;

(7) When operating a hand-held dental intraoral machine, operators shall wear a lead apron as well as an extremity dosimeter on the hand closest to the protective shielding;

(8) A hand-held dental intraoral machine shall be held without any motion during a patient examination. A tube stand or Rinn kit may be utilized to immobilize a hand-held dental intraoral radiographic unit during patient examination;

(9) Operators of hand-held dental intraoral machines shall provide a secondary protective barrier for patients; and

(10) The operator shall ensure that there are no bystanders within a radius of at least 6 feet (2 meters), from the patient being examined with a hand-held intraoral machine.

(b) Additional requirements for operatories in permanent facilities:

(1) Hand-held dental intraoral machines shall be used for patient examinations only in dental operatories that meet the structural shielding requirements specified by DHHS/RHS or by a qualified expert;

(2) Hand-held dental intraoral machines shall not be used for patient examinations in hallways and waiting rooms; and

(3) For battery-powered units, visual means shall be provided on the control panel to indicate whether the battery is in a state of charge adequate for proper operation.

#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07

Disclaimer: These regulations may not be the most recent version. New Hampshire may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.