New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4046 - USE OF RADIATION MACHINES: GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC AND VETERINARY X-RAY SYSTEMS
Section He-P 4046.08 - Dental Intraoral Radiographic Systems
Current through Register No. 40, October 3, 2024
In addition to the provisions of He-P 4045 and He-P 4046.01, the requirements of this section shall apply to x-ray systems used for dental radiography:
(a) X-ray systems designed for use with an intraoral image receptor shall be provided with means to limit SSD, to not less than:
(b) X-ray systems designed for use with an intraoral image receptor shall be provided with means to limit the x-ray beam such that the beam at the minimum SSD shall be containable in a circle having a diameter of no more than 7 centimeters;
(c) The radiation exposure control shall meet the following requirements:
(d) When the equipment is operated on an adequate power supply as specified by the manufacturer, the estimated coefficient of variation of radiation exposures shall be no greater than 0.05, for any specific combination of selected technique factors;
(e) Deviation of technique factors from indicated values for kVp and exposure time shall not exceed the limits specified for that system by its manufacturer or in the absence of manufacturer's specifications the deviation shall not exceed 10 percent of the indicated value for kVp and 20 percent for time;
(f) Dental x-ray systems with a nominal fixed kVp of less than 50 kVp shall not be used to make diagnostic dental radiographs of humans; and
(g) The following administrative controls shall be in place:
#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07