New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4046 - USE OF RADIATION MACHINES: GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC AND VETERINARY X-RAY SYSTEMS
Section He-P 4046.05 - Fluoroscopic X-Ray Systems - Additional Requirements
Current through Register No. 40, October 3, 2024
(a) During fluoroscopy and cine-fluorography the kV and the mA shall be continuously indicated.
(b) The source to skin distance (SSD) shall not be less than:
(c) Fluoroscopic timers shall:
(d) With the exception of C-arm fluoroscopes, fluoroscopic table designs in combination with procedures shall be such that no unprotected part of any staff or ancillary person's body is exposed to unattenuated scattered radiation generated under the table.
(e) The attenuation required by He-P 4046. 05(d) shall not be less than 0.25 millimeter lead equivalent.
(f) With the exception of C-arm fluoroscopes, equipment configuration when combined with procedures shall be such that no portion of any staff or ancillary person's body, except the extremities, is exposed to the unattenuated scattered radiation generated above the tabletop unless either:
(g) Upon receipt of a written request, DHHS/RHS shall grant exemptions to He-P 4046.05(f) where a required sterile field will not permit the use of normal protective barriers for any fluoroscopic procedures.
(h) Fluoroscopic systems equipped with spot-film mode shall meet the exposure reproducibility requirements of He-P 4046 when operating in the spot-film mode.
(i) Radiation therapy simulation systems shall be exempt from:
(j) Procedures for radiation therapy simulation systems shall require that the timer be reset between treatments.
(k) All fluoroscopic equipment displays manufactured on or after June 10, 2006 shall:
#6827, eff 8-6-98; ss by #8692, INTERIM, eff 7-27-06, EXPIRES: 2-2-07; ss by #8808, eff 1-24-07