New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4035 - USE OF RADIONUCLIDES IN THE HEALING ARTS
Section He-P 4035.51 - Safety Precautions for Remote Afterloader Units, Teletherapy Units and Gamma Stereotactic Radiosurgery Units
Current through Register No. 40, October 3, 2024
(a) A licensee shall control access to the treatment room by a door at each entrance.
(b) A licensee shall equip each entrance to the treatment room with an electrical interlock system that shall:
(c) A licensee shall require any individual entering the treatment room to assure, through the use of appropriate radiation monitors, that radiation levels have returned to ambient levels.
(d) Except for low-dose remote afterloader units, a licensee shall construct or equip each treatment room with viewing and intercom systems to permit continuous observation of the patient or the human research subject from the treatment console during irradiation.
(e) For licensed activities where sources are placed within the patient's or human research subject's body, a licensee shall only conduct treatments which allow for expeditious removal of a decoupled or jammed Source.
(f) In addition to the requirements specified in He-P 4035.51(a) through(e), a licensee shall:
(g) A licensee shall have applicable emergency response equipment available near each treatment room to respond to a source:
(See Revision Note at part heading for He-P 4035) #6942, eff 2-1-99; ss by #8800, INTERIM, eff 2-1-07, EXPIRED: 7-31-07
New. #8959, eff 8-7-07