New Hampshire Code of Administrative Rules
He - Department of Health and Human Services
Subtitle He-P - Former Division of Public Health Services
Chapter He-P 4000 - NEW HAMPSHIRE RULES FOR THE CONTROL OF RADIATION
Part He-P 4035 - USE OF RADIONUCLIDES IN THE HEALING ARTS
Section He-P 4035.17 - Possession, Use, Calibration, and Check of Dose Calibrators

Universal Citation: NH Admin Rules He-P 4035.17

Current through Register No. 40, October 3, 2024

(a) For direct measurements performed in accordance with He-P 4035.19, a licensee shall possess and use instrumentation to measure the activity of unsealed byproduct materials before it is administered to each patient or human research subject.

(b) A licensee shall calibrate the instrumentation required in He-P 4035.17(a) in accordance with nationally recognized standards or the manufacturer's instructions.

(c) A licensee shall retain a record of each instrument calibration for 3 years. The records shall include:

(1) The model and serial number of the instrument;

(2) The date of the calibration;

(3) The results of the calibration; and

(4) The name of the individual who performed the calibration.

(See Revision Note at part heading for He-P 4035) #6942, eff 2-1-99; ss by #8800, INTERIM, eff 2-1-07, EXPIRED: 7-31-07

New. #8959, eff 8-7-07

Disclaimer: These regulations may not be the most recent version. New Hampshire may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.