Nevada Administrative Code
Chapter 484C - Testing for Intoxication
EVIDENTIAL BREATH-TESTING DEVICES
Section 484C.130 - Duties of operator

Universal Citation: NV Admin Code 484C.130

Current through December 31, 2024

A certified operator of an evidential breath-testing device:

1. In the ordinary course of his or her business and at the time the operator is performing a test of a person's breath, shall use, follow the instructions in and enter the appropriate information on the Checklist for Operators of Breath-Testing Devices which is approved by the Committee for use with the type of device that he or she is operating.

2. Immediately before performing the evidentiary test of the person's breath, shall verify the calibration of the device by testing it using an aqueous solution or gas which is certified to contain a specific concentration of alcohol within the range that corresponds to a concentration of alcohol in the breath from .05 to .20 gram per 210 liters of breath, inclusive.

(Added to NAC by Com. on Testing for Intoxication, eff. 10-26-83; A 3-11-86; 11-23-93; 9-9-94; R112-97, 5-26-98; R137-99, 12-13-99; R032-05, 10-31-2005) - (Substituted in revision for NAC 484.670)

NRS 484C.620, 484C.640

Disclaimer: These regulations may not be the most recent version. Nevada may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.