Nevada Administrative Code
Chapter 459 - Hazardous Materials
RADIATION CONTROL
X-Rays in the Healing Arts
Section 459.620 - Other radiographic systems: Special purpose systems
Current through December 31, 2024
For special purpose X-ray systems:
1. A means must be provided to limit the X-ray field in the plane of the image receptor so that the X-ray field is within 2 percent of the source-image receptor distance in each dimension of the X-ray field when the axis of the X-ray beam is perpendicular to the plane of the image receptor.
2. A means must be provided to align the center of the X-ray field with the center of the image receptor to within 2 percent of the source-image receptor distance.
3. Subsections 1 and 2 may be met with a system that meets the requirements for a general purpose X-ray system as specified in NAC 459.618, or, when alignment means are also provided, may be met with either:
Bd. of Health, Radiation Control Reg. §§ 6.6.1.4-6.6.1.4.3.2, eff. 2-28-80; A by R144-13A, eff. 10/13/2016
NRS 459.201