Section 459.618 - General purpose X-ray systems: Stationary and mobile
Current through December 31, 2024
1. A means must be provided for stepless adjustment of the size of the X-ray field.
2. A means must be provided for visually defining the perimeter of the X-ray field. The total misalignment of the edges of the visually defined field with the respective edges of the X-ray field along either the length or width of the visually defined field must not exceed 2 percent of the distance from the source to the center of the visually defined field when the surface upon which it appears is perpendicular to the axis of the X-ray beam.
3. The Division may grant an exemption from subsections 1 and 2 for an uncertified X-ray system if the registrant makes a written application for the exemption and in his or her application demonstrates that:
4. All stationary general purpose X-ray systems must meet the following additional requirements:
5. Radiographic equipment designed for only one image receptor size at a fixed source-image receptor distance must be provided with means to limit the field at the plane of the image receptor to dimensions no greater than those of the image receptor, and to align the center of the X-ray field with the center of the image receptor to within 2 percent of the source-image receptor distance.
6. All radiographic equipment must:
Bd. of Health, Radiation Control Reg. §§ 6.6.1.1-6.6.1.3, eff. 2-28-80-NAC A 4-27-84; A by R144-13A, eff. 10/13/2016
NRS 459.201