Section 459.606 - X-ray and electron therapy installations: Focal spot; beam orientation; system checks
Universal Citation: NV Admin Code 459.606
Current through December 31, 2024
1. The registrant must determine or obtain from the manufacturer the location with reference to an accessible point on the radiation head of:
(a) The X-ray target and the virtual source of X rays;
(b) The electron window or the scattering foil, or both; and
(c) All possible orientations of the useful beam.
2. Facilities must be provided so that all radiation safety interlocks can be checked. When preselection of any of the operating conditions requires action in the treatment room and at the treatment control panel, selection at one location must not give a display at the other location until the requisite selected operations in both locations have been completed.
Bd. of Health, Radiation Control Reg. §§ 6.9.1.13-6.9.1.14, eff. 2-28-80
NRS 459.201
Disclaimer: These regulations may not be the most recent version. Nevada may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.