Nevada Administrative Code
Chapter 459 - Hazardous Materials
RADIATION CONTROL
X-Rays in the Healing Arts
Section 459.5936 - Electronic brachytherapy devices: Acceptance test; comparison of position indicators; evaluation and approval of parameters for treatment
Current through December 31, 2024
1. Where applicable, an authorized medical physicist for electronic brachytherapy shall perform an acceptance test on the treatment planning system of computer systems used for therapy, using a published protocol which is accepted by a nationally recognized body. The acceptance test must verify, as applicable:
2. The position indicators in an applicator must be compared to the actual position of the source or planned dwell positions as appropriate at the time of commissioning.
3. Before each regimen for patient treatment, the parameters for the treatment must be evaluated and approved by the authorized user and the authorized medical physicist for electronic brachytherapy for accuracy through means which are independent of those that were used for the determination of the parameters.
Added to NAC by Bd. of Health by R185-08, eff. 5-7-2010
NRS 459.201