Nevada Administrative Code
Chapter 459 - Hazardous Materials
RADIATION CONTROL
X-Rays in the Healing Arts
Section 459.5935 - Electronic brachytherapy devices: Requirements for mobile electronic brachytherapy

Universal Citation: NV Admin Code 459.5935

Current through December 31, 2024

A registrant who provides services for mobile electronic brachytherapy shall:

1. Check all survey instruments before medical use at each location of use or on each day of use, whichever is more frequent;

2. Account for the X-ray tube in the system before departing from a location; and

3. Perform all the periodic spot checks required by NAC 459.5934 at each location.

Added to NAC by Bd. of Health by R185-08, eff. 5-7-2010

NRS 459.201

Disclaimer: These regulations may not be the most recent version. Nevada may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.