Nevada Administrative Code
Chapter 459 - Hazardous Materials
RADIATION CONTROL
X-Rays in the Healing Arts
Section 459.590 - Therapeutic X-ray systems: Requirements for design of treatment room
Current through December 31, 2024
In addition to providing shielding adequate to meet the requirements of NAC 459.320 to 459.374, inclusive, the design of the treatment room must meet the following requirements:
1. Treatment rooms to which access is possible through more than one entrance must be provided with warning lights in a readily observable position near the outside of all access doors, preferably at eye level, which will indicate when the useful beam is on.
2. Provision must be made for two-way aural communication with the patient from the control room.
3. Windows, mirror systems or closed-circuit television viewing screens or equivalent systems must be provided to permit continuous observation of the patient during irradiation and must be so located that the operator may see the patient and the control panel from the same position. When the viewing system is by electronic means, for example, television, an alternate viewing system must be available as a backup in case of electronic failure.
4. Facilities which contain an X-ray system which may be operated above 150 kVp must:
Bd. of Health, Radiation Control Reg. §§ 6.8.2-6.8.2.4.5.1, eff. 2-28-80-NAC A 4-27-84; 1-18-94
NRS 459.201