Nevada Administrative Code
Chapter 459 - Hazardous Materials
RADIATION CONTROL
X-Rays in the Healing Arts
Section 459.588 - Therapeutic X-ray systems: Timer; control panel
Current through December 31, 2024
1. A timer must be provided which has a display at the treatment control panel. The timer must have a preset timer selector and an elapsed time indicator.
2. The timer must be a cumulative timer which switches on and off with the radiation and retains its reading after irradiation is interrupted or terminated. It must be necessary to zero the elapsed time indicator and the preset time selector after irradiation is terminated.
3. The timer must terminate irradiation when a preselected time has elapsed.
4. The control panel must have:
5. When a control panel may energize more than one X-ray tube:
6. There must be means of determining the target to patient distance to within 1 centimeter.
7. Unless it is possible to bring the X-ray output to the prescribed exposure parameters within 5 seconds, the entire useful beam must be attenuated by a shutter having a lead equivalency not less than that of the tube housing. In addition:
Bd. of Health, Radiation Control Reg. §§ 6.8.1.9-6.8.1.13.2, eff. 2-28-80
NRS 459.201