Nevada Administrative Code
Chapter 459 - Hazardous Materials
RADIATION CONTROL
X-Rays in the Healing Arts
Section 459.576 - Fluoroscopic X-ray systems: Fluoroscopic timer; mobile fluoroscopes
Universal Citation: NV Admin Code 459.576
Current through December 31, 2024
1. A means must be provided to preset the cumulative on-time of the fluoroscopic tube. The maximum cumulative time of the timing device must not exceed 5 minutes without resetting.
2. A signal audible to the fluoroscopist must indicate the completion of any preset cumulative on-time. Such a signal must continue to sound while X rays are produced until the timing device is reset. In addition to the other requirements of NAC 459.566 to 459.578, inclusive, mobile fluoroscopes must provide image intensification.
Bd. of Health, Radiation Control Reg. §§ 6.5.7-6.5.8, eff. 2-28-80
NRS 459.201
Disclaimer: These regulations may not be the most recent version. Nevada may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.