Nevada Administrative Code
Chapter 457 - Cancer
OPERATION OF RADIATION MACHINE FOR MAMMOGRAPHY
Tests for Quality Assurance
Section 457.400 - Tests required after certain repairs or replacements of components

Universal Citation: NV Admin Code 457.400

Current through December 31, 2024

Tests for quality assurance must be performed pursuant to 21 C.F.R. § 900.12 whenever any component of the radiation machine is repaired or replaced, and before the machine is used for mammography if the repair or replacement affects:

1. Image quality.

2. The accuracy or consistency of operation of the exposure timer or automatic exposure control.

3. Milliampere-seconds linearity.

4. The accuracy of the kilovolt peak indicated by the machine.

5. The skin entrance exposure or glandular tissue dose.

6. Focal spot size.

7. Half-value layer.

Added to NAC by Bd. of Health, eff. 1-24-92; A 11-1-95; R033-06, 6-28-2006; A by R144-13A, eff. 10/13/2016

NRS 457.065

Disclaimer: These regulations may not be the most recent version. Nevada may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.