Nevada Administrative Code
Chapter 444 - Sanitation
INVASIVE BODY DECORATION ESTABLISHMENTS
Maintenance of Facilities, Equipment and Supplies
Section 444.00911 - Requirements for sterilizer; check of expiration date before use

Universal Citation: NV Admin Code 444.00911

Current through September 16, 2024

1. A sterilizer used in an invasive body decoration establishment must:

(a) Be manufactured for the sterilization of medical instruments;

(b) Meet the specifications set forth in the standards adopted by reference in paragraph (a) of subsection 1 of section 35;

(c) Be tested using a commercial biological indicator monitoring system:
(1) After the initial installation;

(2) After any service or repair; and

(3) Once each month; and

(d) Be loaded, operated, disinfected and maintained according to the manufacturer's specifications

2. Before each use, the operator of the sterilizer must check the expiration date of the biological indicator monitoring system.

Added to NAC by Bd. of Health by R112-16A, eff. 12/19/2017

NRS 439.200

Disclaimer: These regulations may not be the most recent version. Nevada may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.