A licensee must retain a record of each spot-check for remote afterloader units required by 180 NAC 7-077 for three years. The record must include, as applicable:

(A) The date of the spot-check;

(B) The manufacturer's name, model number, and serial number for the remote afterloader unit and source;

(C) An assessment of timer accuracy;

(D) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and

(E) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.

Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.