Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 180 - CONTROL OF RADIATION
Chapter 7 - MEDICAL USE OF RADIOACTIVE MATERIAL
Section 180-7-107 - RECORDS OF DOSIMETRY EQUIPMENT
Current through September 17, 2024
A licensee must retain a record of the calibration, intercomparison, and comparisons of its dosimetry equipment done according to 180 NAC 7-072 for the duration of the license. For each calibration, intercomparison, or comparison, the record must include:
(A) The date;
(B) The manufacturer's name, model numbers and serial numbers of the instruments that were calibrated, intercompared, or compared as required by 180 NAC 7-072.01 and 7-072.02;
(C) The correction factor that was determined from the calibration or comparison or the apparent correction factor that was determined from an intercomparison; and
(D) The names of the individuals who performed the calibration, intercomparison, or comparison.