Section 180-7-072 - DOSIMETRY EQUIPMENT
Current through September 17, 2024
072.01 CALIBRATED DOSIMETRY SYSTEM. Other than LDR remote afterloader sources where the source output or activity is determined by the manufacturer, a licensee must have a calibrated dosimetry system available for use. To satisfy this requirement, one of the following conditions must be met.
072.02 SPOT-CHECK SYSTEM. The licensee must have available for use a dosimetry system for spot-check output measurements, if applicable. To satisfy this requirement, the system may be compared with a system calibrated according to 180 NAC 7-072.01. This comparison must have been performed within the previous year and after each servicing that may have affected system calibration. The spot-check system may be the same system used to meet the requirement in 180 NAC 7-072.01.
072.03 RECORDS. The licensee must retain a record of each calibration, intercomparison, and comparison according to 180 NAC 7-107.