Section 180-7-061 - THERAPY-RELATED COMPUTER SYSTEMS

Current through September 17, 2024

The licensee must perform acceptance testing on the treatment planning system of therapy-related computer systems according to published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:

(A) The source-specific input parameters required by the dose calculation algorithm;

(B) The accuracy of dose, dwell time, and treatment time calculations at representative points;

(C) The accuracy of isodose plots and graphic displays; and

(D) The accuracy of the software used to determine sealed source positions from radiographic images.

Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.