Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 180 - CONTROL OF RADIATION
Chapter 7 - MEDICAL USE OF RADIOACTIVE MATERIAL
Section 180-7-060 - CALIBRATION MEASUREMENTS OF BRACHYTHERAPY SOURCES
Current through September 17, 2024
060.01 USE PARAMETERS. Before the first medical use of a brachytherapy source a licensee must:
060.02 SOURCE MANUFACTURER OR CALIBRATION LABORATORY. A licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine that are made according to 180 NAC 7-060.01.
060.03 PHYSICAL DECAY CORRECTION. A licensee must mathematically correct the outputs or activities determined in 180 NAC 7-060.01 for physical decay at intervals consistent with one percent physical decay.
060.04 PERFORM OR REVIEW OF CALCULATION MEASUREMENTS. An authorized medical physicist must perform or review the calculation measurements made according to 180 NAC 7-060.01, 7-060.02, or 7-060.03.
060.05 STRONTIUM-90 SOURCES FOR OPHTHALMIC TREATMENTS. Requirements for licensees using strontium-90 sources for ophthalmic treatments are as follows:
060.06 RECORDS. A licensee must retain a record of each calibration according to 180 NAC 7-104.