Section 180-7-029 - POSSESSION, USE, AND CALIBRATION OF INSTRUMENTS USED TO MEASURE THE ACTIVITY OF UNSEALED RADIOACTIVE MATERIAL
Universal Citation: 180 NE Admin Rules and Regs ch 7 ยง 029
Current through September 17, 2024
To perform direct measurements according to 180 NAC 7-031, a licensee must:
(A) Possess and use instrumentation to measure the activity of unsealed radioactive material before administration to each patient or human research subject;
(B) Calibrate the instrumentation according to nationally recognized standards or the manufacturer's instructions. The calibrations must, at minimum and as appropriate to demonstrate proper operation of the instrument, test for:
(i)
Constancy;
(ii)
Linearity;
(iii) Accuracy;
and
(iv) Geometry dependence;
and
(C) Retain a record of each instrument test required by 180 NAC 7-029 according to 180 NAC 7-091.
Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.