Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 180 - CONTROL OF RADIATION
Chapter 6 - DIAGNOSTIC X RAYS IN THE HEALING ARTS
Section 180-6-005 - FLUOROSCOPIC X-RAY SYSTEMS

Current through March 20, 2024

Fluoroscopic x-ray systems must meet the machine performance standards of 21 CFR § 1020.32. Use of non-image intensified fluoroscopic equipment is prohibited. The provisions of this chapter apply to equipment for fluoroscopic imaging or for recording images from the fluoroscopic image receptor, except computed tomography (CT) x-ray systems manufactured on or after November 29, 1984.

005.01 PERIODIC MEASUREMENT OF AIR KERMA RATE (AKR). A periodic measurement of air kerma rate (AKR) must be performed as follows.

005.01(A) MEASUREMENTS OF AIR KERMA RATE (AKR). Measurements must be made for both typical and maximum values of the air kerma rate (AKR), materials may be placed in the useful beam to protect the imaging system when performing the periodic measurements. 005.01(A)(i) Measurements must be made annually or after any maintenance of the system which might affect the air kerma rate (AKR. 005.01(A)(ii) For units manufactured before June 10, 2006, results of the measurements must be posted where a fluoroscopist may have access to those results while using the fluoroscope.

005.01(B) RECORDS OF MEASUREMENTS OF AIR KERMA RATE (AKR).The measurement results may be stated in roentgens per minute (R/min) or milliGray per minute (mGy/min). The results must include the technique factors used to determining the results, the name of the individual that performed the measurements, and the date the measurements were performed.

005.01(C) CONDITIONS OF PERIODIC MEASUREMENT OF TYPICAL AIR KERMA RATE (AKR). The following conditions apply to the periodic measurement of the typical Air Kerma Rate (AKR):
(i) The measurement must be made under the conditions that satisfy the requirements of 21 CFR § 1020.32(d)(3);

(ii) Fluoroscopic systems that do not incorporate automatic exposure rate control (AERC) must use a milliamperage (mA) and kilovolt peak (kVp) typical of clinical use of the fluoroscopic system; and

(iii) Fluoroscopic systems that incorporate automatic exposure rate control (AERC) must have sufficient material placed in the useful beam to produce a milliamperage (mA) and kilovolt peak (kVp) typical of the clinical use of the fluoroscopic system.

005.01(D) CONDITIONS OF PERIODIC MEASUREMENT OF MAXIMUM AIR KERMA RATE (AKR). The following conditions apply to the periodic measurement of the maximum air kerma rate (AKR):
(i) The measurement must be made under the conditions that satisfy the requirements of 21 CFR § 1020.32(d)(3);

(ii) Fluoroscopic systems that do not incorporate automatic exposure rate control (AERC) must be adjusted to those settings which give the maximum air kerma rate (AKR); and

(iii) Fluoroscopic systems that incorporate automatic exposure rate control (AERC) must have sufficient material placed in the useful beam to produce the maximum air kerma rate (AKR) of the system.

005.02 CONTROL OF SCATTERED RADIATION. The following requirements apply to controlling the scatter of radiation.

005.02(A) FLUOROSCOPIC TABLE DESIGNS. Fluoroscopic table designs, when combined with the procedures performed at the registrant's facility, must ensure that no unprotected part of any staff or ancillary individual's body is exposed to unattenuated scattered radiation that originates from under the table. The attenuation provided must be not less than 0.25 millimeter lead equivalent.

005.02(B) EQUIPMENT CONFIGURATION. Equipment configuration, when combined with procedures performed at the registrant's facility, must ensure that no portion of any staff or ancillary individual's body, except the extremities, is exposed to the unattenuated scattered radiation that originates from above the tabletop unless:
(i) That individual is 120 centimeters from the center of the useful beam; or

(ii) The radiation has passed through not less than 0.25 millimeter lead equivalent material including, but not limited to, drapes, bucky-slot cover panel, or self-supporting curtains. This requirement is in addition to any lead equivalency provided by the protective apron specified in 180 NAC 003.08.

005.02(C) EXEMPTIONS TO THE USE OF PROTECTIVE BARRIERS. When a sterile field will not permit the use of the normal protective barriers or drapes, the shielding required by 180 NAC 6-005.02(B) must be maintained to the degree possible under the clinical conditions.

005.03 FLUOROSCOPIC RADIATION THERAPY SIMULATION SYSTEMS. Fluoroscopic radiation therapy simulation systems are exempt from the requirements of 180 NAC 6-005.01. In addition, these systems are exempt from the requirements of 21 CFR § 1020.32(a) provided these systems are designed and used in a manner that no individual other than the patient is in the x-ray room during the time the system is producing x-rays.

005.04 EQUIPMENT OPERATION. The following requirements apply to the operation of fluoroscopic equipment.

005.04(A) Images formed by the use of fluoroscopic x-ray systems must be under the direction of and interpreted by a licensed practitioner of the healing arts.

005.04(B) Only a licensed practitioner may perform interpretative fluoroscopic procedures.

005.04(C) Fluoroscopy must not be used as a positioning tool for general purpose radiographic examinations.

005.04(D) Operators must be competent in the standard operating procedures of the unit in use.

005.04(E) Registrants must maintain a record of the cumulative fluoroscopic exposure time used and the number of fluorographic images recorded for each examination. This record must include patient identification, type and date of examination, the fluoroscopic system used, and operator's name.

Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.