Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 180 - CONTROL OF RADIATION
Chapter 6 - DIAGNOSTIC X RAYS IN THE HEALING ARTS
Section 180-6-004 - GENERAL REQUIREMENTS FOR ALL DIAGNOSTIC AND INTERVENTIONAL X-RAY SYSTEMS
Current through September 17, 2024
In addition to other requirements of 180 NAC 6-004, all diagnostic and interventional x-ray systems must meet the specifications of 21 CFR § 1020.30.
004.01 FILTRATION CONTROL. X-ray systems that have variable kilovolt peak (kVp) and variable filtration for the useful beam must have a device to link the kilovolt peak (kVp) selector with the filter or filters and must prevent an exposure unless the minimum amount of filtration required in 21 CFR § 1020.30(m) is in the useful beam for the kVp that has been selected.
004.02 BEAM LIMITATION. The useful beam must be limited to the area of clinical interest.
004.03 MULTIPLE TUBES. When two or more radiographic tubes are controlled by one exposure switch, the tube or tubes which have been selected must be clearly indicated prior to initiation of the exposure. This indication must be on the x-ray control panel and at or near the tube housing assembly that has been selected.
004.04 MECHANICAL SUPPORT OF TUBE HEAD. The tube housing assembly supports must be adjusted so that the tube housing assembly will remain stable during the exposure except when the tube housing movement is a designed function of the x-ray system.
004.05 MAINTAINING COMPLIANCE. Diagnostic x-ray systems and their associated components used on humans and certified under the Federal X-ray Equipment Performance Standard, 21 CFR Part 1020, must be maintained in compliance with applicable requirements of that standard.
004.06 LOCKS. All position locking, holding, and centering devices on x-ray systems components and systems must function as intended.
004.07 EQUIPMENT PERFORMANCE EVALUATION. For all radiation generating equipment, except bone densitometry , veterinary, computed tomography (CT), and cone beam computed tomography (CBCT) the registrant must perform, or cause to be performed, tests necessary to insure the proper function of equipment. These tests must be performed every three years. For dental radiation generating equipment, these tests must be performed every five years. The evaluation must include the following measures.