Section 180-5-019 - PERSONNEL MONITORING

Current through March 20, 2024

019.01 DIRECT READING DOSIMETER, ALARMING RATE METER AND PERSONNEL DOSIMETER. The licensee or registrant may not permit any individual to act as a radiographer or a radiographer's assistant unless, at all times during radiographic operations, each individual wears, on the trunk of the body, a direct reading dosimeter, an operating alarming ratemeter, and a personnel dosimeter. At permanent radiography installations where other appropriate alarming or warning devices are in routine use, the use of an alarming ratemeter is not required.

(A) Pocket dosimeters must have a range from zero to 2 millisieverts (200 mrem) and must be recharged at the start of each shift. Electronic personal dosimeters may only be used in place of ion-chamber pocket dosimeters.

(B) Each personnel dosimeter must be assigned to and worn by only one individual;

(C) Film badges must be replaced at least monthly and all other personnel dosimeters that require replacement must be replaced at least quarterly. All personnel dosimeters must be evaluated at least quarterly or promptly after replacement, whichever is more frequent.

019.02 BEGINNING AND END OF SHIFT. Direct reading dosimeters such as pocket dosimeters or electronic personal dosimeters, must be read and the exposures recorded at the beginning and end of each shift, and records must be maintained as required by 180 NAC 5-033.

019.03 RESPONSE TO RADIATION CHECK. Pocket dosimeters, or electronic personal dosimeters, must be checked at periods not to exceed 12 months for correct response to radiation, and records must be maintained as required by 180 NAC 5-033. Acceptable dosimeters must read within plus or minus 20% of the true radiation exposure.

019.04 PERSONNEL DOSIMETER PROCESSING. If an individual's pocket chamber is found to be off-scale, or if the individual's electronic personal dosimeter reads greater than 2 millisieverts (200 mrem), and the possibility of radiation exposure cannot be ruled out as the cause, the individual's personnel dosimeter must be sent for processing and evaluation within 24 hours. For personnel dosimeters that do not require processing, evaluation of the dosimeter must be started within 24 hours. In addition, the individual may not resume work associated with licensed material use until a determination of the individual's radiation dose has been made. This determination must be made by the radiation safety officer or the radiation safety officer's designee. The results of this determination must be included in the records maintained as required by 180 NAC 5-033.

019.05 LOST OR DAMAGED PERSONNEL DOSIMETER. If the personnel dosimeter that is required by 180 NAC 5-019 is lost or damaged, the worker must cease work immediately until a replacement personnel dosimeter meeting the requirements of 180 NAC 5-019 is provided and the exposure is calculated for the time period from issuance to loss or damage of the personnel dosimeter. The results of the calculated exposure and the time period for which the personnel dosimeter was lost or damaged must be included in the records maintained as required by 180 NAC 5-033.

019.06 RECORD RETENTION. Dosimetry results must be retained as required by 180 NAC 5-033.

019.07 ALARMING RATEMETER. Each alarming ratemeter must:

(A) Be checked to ensure that the alarm functions properly before using at the start of each shift;

(B) Be set to give an alarm signal at a preset dose rate of 5 millisieverts per hour (500 mrem/hr); with an accuracy of plus or minus 20% of the true radiation dose rate;

(C) Require special means to change the preset alarm function; and

(D) Be calibrated at periods not to exceed 12 months for correct response to radiation. The licensee must maintain records of alarming ratemeter calibrations as required by 180 NAC 5-033.

Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.