Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 180 - CONTROL OF RADIATION
Chapter 5 - RADIATION SAFETY REQUIREMENTS FOR INDUSTRIAL RADIOGRAPHIC OPERATIONS
Section 180-5-008 - RADIATION SURVEY INSTRUMENTS

Current through March 20, 2024

008.01 INSTRUMENT PERFORMANCE REQUIREMENTS. The licensee or registrant must keep sufficient calibrated and operable radiation survey instruments at each location where sources of radiation are present to make the radiation surveys required by 180 NAC 4 and 180 NAC 5. Instrumentation required by 180 NAC 5 must be capable of measuring a range from 0.02 millisieverts (2 mrem) per hour through 0.01 sievert (1 rem) per hour.

008.02 CALIBRATION REQUIREMENTS. The licensee or registrant must have each radiation survey instrument required under 180 NAC 5-008.01 calibrated:

(A) At energies appropriate for use and at intervals not to exceed six months or after instrument servicing, except for battery changes;

(B) For linear scale instruments, at two points located approximately one-third and two-thirds of full-scale on each scale; for logarithmic scale instruments, at mid-range of each decade, and at two points of at least one decade; and for digital instruments, at 3 points between 0.02 and 10 millisieverts (2 and 1000 mrem) per hour; and

(C) So that an accuracy within plus or minus 20% of the true radiation dose rate can be demonstrated at each point checked.

008.03 RECORDS. The licensee or registrant must maintain records of the results of the instrument calibrations as required by 180 NAC 5-025.

Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.