Current through March 20, 2024
007.01
LEAKAGE RADIATION OUTSIDE
THE MAXIMUM USEFUL BEAM IN PHOTON AND ELECTRON MODES. The
requirements governing leakage radiation outside the maximum useful beam in
photon and electron modes are as follows.
007.01(A)
ABSORBED
DOSE. The absorbed dose due to leakage radiation, excluding
neutrons, at any point outside the maximum sized useful beam, but within a
circular plane of radius 2 meters which is perpendicular to and centered on the
central axis of the useful beam at the nominal treatment distance, must not
exceed a maximum of 0.2 percent and an average of 0.1 percent of the absorbed
dose on the central axis of the beam at the nominal treatment distance.
Measurements must be averaged over an area not exceeding 100 square centimeters
at a minimum of 16 points uniformly distributed in the plane.
007.01(B)
LEAKAGE
RADIATION. Except for the area defined in 180 NAC 20-007.01(A),
the absorbed dose due to leakage radiation, excluding neutrons, at 1 meter from
the electron path between the electron source and the target or electron window
must not exceed 0.5 percent of the absorbed dose on the central axis of the
beam at the nominal treatment distance. Measurements must be averaged over an
area not exceeding 100 square centimeters.
007.01(C)
NEUTRON ABSORBED
DOSE. For equipment manufactured after July 11, 2009, the neutron
absorbed dose outside the useful beam must not exceed manufacturer's
specifications.
007.01(D)
LEAKAGE RADIATION MEASUREMENTS. For each therapeutic
radiation machine, the registrant must determine, or obtain from the
manufacturer, the leakage radiation existing at the positions specified in 180
NAC
20-007.01(A) through
(C) for the specified operating conditions.
Records on leakage radiation measurements must be maintained for inspection by
the Department.
007.02
LEAKAGE RADIATION THROUGH BEAM LIMITING DEVICES. The
requirements governing leakage radiation through beam limiting devices are as
follows.
007.02(A)
PHOTON
RADIATION. All adjustable or interchangeable beam limiting devices
must attenuate the useful beam so that at the nominal treatment distance, the
maximum absorbed dose anywhere in the area shielded by the beam limiting device
or devices must not exceed 2 percent of the maximum absorbed dose on the
central axis of the useful beam measured in a 100 square centimeters radiation
field, or maximum available field size if less than 100 square
centimeters.
007.02(B)
ELECTRON RADIATION. All adjustable or interchangeable
electron applicators must attenuate the radiation, including, but not limited
to, photon radiation generated by electrons incident on the beam limiting
device and electron applicator and other parts of the radiation head, so that
the absorbed dose in a plane perpendicular to the central axis of the useful
beam at the nominal treatment distance must not exceed:
(i) A maximum of 2 percent and average of 0.5
percent of the absorbed dose on the central axis of the useful beam at the
nominal treatment distance. This limit must apply beyond a line 7 centimeters
outside the periphery of the useful beam; and
(ii) A maximum of 10 percent of the absorbed
dose on the central axis of the useful beam at the nominal treatment distance.
This limit must apply beyond a line 2 centimeters outside the periphery of the
useful beam.
007.03
MEASUREMENT OF LEAKAGE
RADIATION. The requirements for measurement of leakage radiation
are as follows.
007.03(A)
PHOTON
RADIATION. Measurements of leakage radiation through the beam
limiting devices must be made with the beam limiting devices closed and any
residual aperture blocked by a minimum of two tenth value layers of suitable
absorbing material. In the case of overlapping beam limiting devices, the
leakage radiation through each set must be measured independently at the depth
of maximum dose. Measurements must be made, using a radiation detector, of an
area not exceeding 10 square centimeters.
007.03(B)
ELECTRON
RADIATION. Measurements of leakage radiation through the electron
applicators must be made with the electron beam directed into the air and using
a radiation detector of an area up to, but not exceeding, one square centimeter
suitably protected against radiation that has been scattered from material
beyond the radiation detector. Measurements must be made using one centimeter
of water equivalent build up material.
007.04
FILTERS AND
WEDGES. The requirements governing filters and wedges are as
follows.
007.04(A)
WEDGE
FILTERS. Each wedge filter that is removable from the system must
be clearly marked with an identification number. For removable wedge filters,
the nominal wedge angle must appear on the wedge, or wedge tray if permanently
mounted to the tray. If the wedge or wedge tray is significantly damaged, the
wedge transmission factor must be redetermined.
007.04(B)
ABSORBED
DOSE. If the absorbed dose rate information required by 180 NAC
20-007.01 relates exclusively to
operation with a field flattening filter or beam scattering foil in place, such
foil or filter must be removable only by the use of tools.
007.04(C)
WEDGE FILTERS,
INTERCHANGEABLE FIELD FLATTENING FILTERS, OR INTERCHANGEABLE BEAM SCATTERING
FOILS. For equipment installed after July 11, 2009 that utilizes
wedge filters, interchangeable field flattening filters, or interchangeable
beam scattering foils, the following apply:
(i) Irradiation must not be possible until a
selection of a filter, or a selection to use "no filter", has been made at the
treatment control panel, either manually or automatically;
(ii) An interlock system must be provided to
prevent irradiation if the filter selected is not in the correct
position;
(iii) A display must be
provided at the treatment control panel showing the wedge filter or filters,
interchangeable field flattening filter or filters, or interchangeable beam
scattering foil or foils in use; and
(iv) An interlock must be provided to prevent
irradiation if any filter or beam scattering foil selection operation carried
out in the treatment room does not agree with the filter or beam scattering
foil selection operation carried out at the treatment control panel.
007.05
STRAY
RADIATION IN THE USEFUL BEAM. For equipment installed after July
11, 2009, the registrant must determine during acceptance testing, or obtain
from the manufacturer, data sufficient to ensure that x-ray stray radiation in
the useful electron beam, absorbed dose at the surface during x-ray
irradiation, and stray neutron radiation in the useful x-ray beam do not exceed
manufacturer's specifications.
007.06
BEAM
MONITORS. All therapeutic radiation machines subject to 180 NAC
20-007 must be provided with redundant beam monitoring systems. The sensors for
these systems must be fixed in the useful beam during treatment to indicate the
dose monitor unit rate.
007.06(A)
INDEPENDENTLY POWERED INTEGRATING DOSE METERS.
Equipment installed after July 11, 2009 must be provided with at least two
independently powered integrating dose meters. Alternatively, common elements
may be used if the production of radiation is terminated upon failure of any
common element.
007.06(B)
RADIATION DETECTOR. Equipment installed on or before
July 11, 2009, must be provided with at least one radiation detector. This
detector must be incorporated into a useful beam monitoring system;
007.06(C)
DETECTOR AND
SYSTEM. The detector and the system into which that detector is
incorporated must meet the following requirements:
(i) Each detector must be removable only with
tools and, if movable, must be interlocked to prevent incorrect
positioning;
(ii) Each detector
must form part of a beam monitoring system from whose readings in dose monitor
units the absorbed dose at a reference point can be calculated;
(iii) Each beam monitoring system must be
capable of independently monitoring, interrupting, and terminating
irradiation;
(iv) For equipment
installed after July 11, 2009, the design of the beam monitoring systems must
ensure that the:
(1) Malfunctioning of one
system must not affect the correct functioning of the other system or systems;
and
(2) Failure of either system
must terminate irradiation or prevent the initiation of radiation;
and
(v) Each beam
monitoring system must have a legible display at the treatment control panel.
For equipment installed after July 11, 2009, each display must:
(1) Maintain a reading until intentionally
reset;
(2) Have only one scale and
no electrical or mechanical scale multiplying factors;
(3) Utilize a design so that increasing dose
is displayed by increasing numbers; and
(4) In the event of power failure, the beam
monitoring information required in 180 NAC 20-007.06(C)(v)(3) displayed at the
control panel at the time of failure must be retrievable in at least one system
for a 20 minute period of time.
007.07
BEAM
SYMMETRY. The requirements for beam symmetry are as follows.
007.07(A)
AUXILIARY
DEVICES. Bent beam linear accelerators with a beam flattening
filter or filters subject to 180 NAC 20-007 must be provided with an auxiliary
device or devices to monitor beam symmetry.
007.07(B)
DETECTION OF
ASYMMETRY. The device or devices referenced in 180 NAC
20-007.07(A) must
be able to detect field asymmetry greater than 10 percent.
007.07(C)
TERMINATE
IRRADIATION. The device or devices referenced in 180 NAC
20-007.07(A) must
be configured to terminate irradiation if the specifications in 180 NAC
20-007.07(B)
cannot be maintained.
007.08
SELECTION AND DISPLAY OF
DOSE MONITOR UNITS. The requirements for selection and display of
dose monitor units are as follows.
007.08(A)
SELECTION OF DOSE MONITORING UNITS. Irradiation must
not be possible until a new selection of a number of dose monitor units has
been made at the treatment control panel.
007.08(B)
DISPLAY OF DOSE
MONITORING UNITS. The pre-selected number of dose monitor units
must be displayed at the treatment control panel until reset manually for the
next irradiation.
007.08(C)
RESETTING OF DOSIMETER DISPLAY. After termination of
irradiation, it must be necessary to reset the dosimeter display before
subsequent treatment can be initiated.
007.08(D)
RESET OF PRE-SELECTED
DOSE MONITOR UNITS. For equipment installed after July 11, 2009
after termination of irradiation, it must be necessary for the operator to
reset the pre-selected dose monitor units before irradiation can be
initiated.
007.09
AIR KERMA RATE OR ABSORBED DOSE RATE. For equipment
installed after July 11, 2009, a system must be provided from where the
readings of the air kerma rate or absorbed dose rate at a reference point can
be calculated. The radiation detectors specified in 180 NAC
20-007.06 may form part of this
system. In addition:
(A) The dose monitor
unit rate must be displayed at the treatment control panel;
(B) If the equipment can deliver, under any
conditions, an air kerma rate or absorbed dose rate at the nominal treatment
distance more than twice the maximum value specified by the manufacturer, a
device must be provided which terminates irradiation when the air kerma rate or
absorbed dose rate exceeds a value twice the specified maximum. A record of the
dose rate that the irradiation will be terminated must be maintained by the
registrant;
(C) If the equipment
can deliver, under any fault condition or conditions, an air kerma rate or
absorbed dose rate at the nominal treatment distance more than ten times the
maximum value specified by the manufacturer, a device must be provided to
prevent the air kerma rate or absorbed dose rate anywhere in the radiation
field from exceeding twice the specified maximum value and to terminate
irradiation if the excess absorbed dose at the nominal treatment distance
exceeds 4 gray (Gy) or 400 rad; and
(D) For each therapeutic radiation machine
the registrant must determine, or obtain from the manufacturer, the maximum
value or values specified in 180 NAC
20-007.09(B) and
(C), for the specified operating conditions.
Records of these maximum value or values must be maintained for inspection by
the Department.
007.10
TERMINATION OF IRRADIATION BY THE BEAM MONITORING SYSTEM OR SYSTEMS
DURING STATIONARY BEAM RADIATION THERAPY. The requirements for
termination of irradiation by the beam monitoring system or systems during
stationary beam radiation therapy are as follows.
007.10(A)
PRE-SELECTED NUMBER OF
DOSE MONITOR UNITS. Each primary system must terminate irradiation
when the pre-selected number of dose monitor units has been detected by the
system.
007.10(B)
SECONDARY DOSE MONITORING SYSTEM. If the original
design of the equipment included a secondary dose monitoring system, that
system must be capable of terminating irradiation when not more than 15 percent
or 40 dose monitor units above the pre-selected number of dose monitor units
set at the control panel has been detected by the secondary dose monitoring
system.
007.10(C)
CONTROL PANEL INDICATION. For equipment installed
after July 11, 2009, an indicator on the control panel must show which
monitoring system has terminated irradiation.
007.11
TERMINATION OF
IRRADIATION. It must be possible to terminate irradiation and
equipment movement or go from an interruption condition to termination
condition at any time from the operator's position at the treatment control
panel.
007.12
INTERRUPTION OF IRRADIATION. If a therapeutic
radiation machine has an interrupt mode, it must be possible to interrupt
irradiation and equipment movements at any time from the treatment control
panel. Following an interruption it must be possible to restart irradiation by
operator action without any reselection of operating conditions. If any change
is made of a pre-selected value during an interruption, irradiation and
equipment movements must be automatically terminated.
007.13
TIMER. A
suitable irradiation control device must be provided to terminate the
irradiation after a pre-set time interval as follows:
(A) A timer must be provided which has a
display at the treatment control panel. The timer must have a pre-set time
selector and an elapsed time indicator;
(B) The timer must be a cumulative timer that
activates with an indication of "BEAM-ON" and retains its reading after
irradiation is interrupted or terminated. After irradiation isterminated and
before irradiation can be reinitiated, it must be necessary to reset the
elapsed time indicator; and
(C) The
timer must terminate irradiation when a pre-selected time has elapsed, if the
dose monitoring systems have not previously terminated
irradiation.
007.14
SELECTION OF RADIATION TYPE. Equipment capable of both
x-ray therapy and electron therapy must meet the following additional
requirements:
(A) Irradiation must not be
possible until a selection of radiation type, x-rays or electrons, has been
made at the treatment control panel;
(B) The radiation type selected must be
displayed at the treatment control panel before and during
irradiation;
(C) An interlock
system must be provided to ensure that the equipment can principally emit only
the radiation type that has been selected;
(D) An interlock system must be provided to
prevent irradiation with x-rays, except to obtain an image, when electron
applicators are fitted;
(E) An
interlock system must be provided to prevent irradiation with electrons when
accessories specific for x-ray therapy are fitted; and
(F) An interlock system must be provided to
prevent irradiation if any selected operations carried out in the treatment
room do not agree with the selected operations carried out at the treatment
control panel.
007.15
SELECTION OF ENERGY. Equipment capable of generating
radiation beams of different energies must meet the following requirements:
(A) Irradiation must not be possible until a
selection of energy has been made at the treatment control panel;
(B) The nominal energy value selected must be
displayed at the treatment control panel until reset manually for the next
irradiation. After termination of irradiation, it must be necessary to reset
the nominal energy value selected before subsequent treatment can be
initiated;
(C) Irradiation must not
be possible until the appropriate flattening filter or scattering foil for the
selected energy is in its proper location; and
007.16
SELECTION OF STATIONARY
BEAM RADIATION THERAPY OR MOVING BEAM RADIATION THERAPY.
Therapeutic radiation machines capable of both stationary beam radiation
therapy and moving beam radiation therapy must meet the following requirements:
(A) Irradiation must not be possible until a
selection of stationary beam radiation therapy or moving beam radiation therapy
has been made at the treatment control panel;
(B) The mode of operation must be displayed
at the treatment control panel;
(C)
An interlock system must be provided to ensure that the equipment can operate
only in the mode that has been selected;
(D) An interlock system must be provided to
prevent irradiation if any selected parameter in the treatment room does not
agree with the selected parameter at the treatment control panel;
(E) Moving beam radiation therapy must be
controlled to obtain the selected relationships between incremental dose
monitor units and incremental movement. For equipment installed after July 11,
2009:
(i) An interlock system must be
provided to terminate irradiation if the number of dose monitor units delivered
in any 10 degrees of rotation or 1 centimeter of linear motion differs by more
than 20 percent from the selected value;
(ii) Where the angle terminates the
irradiation in moving beam radiation therapy, the dose monitor units delivered
must differ by less than 5 percent from the dose monitor unit value
selected;
(iii) An interlock must
be provided to prevent motion of more than 5 degrees or 1 centimeter beyond the
selected limits during moving beam radiation therapy;
(iv) An interlock must be provided to require
that a selection of direction be made at the treatment control panel in all
units that are capable of both clockwise and counter clockwise moving beam
radiation therapy; and
(v) Moving
beam radiation therapy must be controlled with both primary position sensors
and secondary position sensors to obtain the selected relationships between
incremental dose monitor units and incremental movement;
(F) Where the beam monitor system terminates
the irradiation in moving beam radiation therapy, the termination of
irradiation must be as required by 180 NAC 20-007.10; and
(G) For equipment installed after July 11,
2009, an interlock system must be provided to terminate irradiation if
movement:
(i) Occurs during stationary beam
radiation therapy; or
(ii) Does not
start or stop during moving beam radiation therapy unless such stoppage is a
pre-planned function.
007.17
FACILITY DESIGN
REQUIREMENTS FOR THERAPEUTIC RADIATION MACHINES OPERATING ABOVE 500 kilovolts
(kV). In addition to shielding adequate to meet the requirements
of 180 NAC
20-009, the facility design must
include the following:
(A) All protective
barriers must be fixed, except for access doors to the treatment room or
movable beam interceptors;
(B) In
addition to other requirements specified in this chapter, the control panel
must also:
(i) Be located outside the
treatment room;
(ii) Provide an
indication of whether electrical power is available at the control paneland if
activation of the radiation is possible;
(iii) Provide an indication of whether
radiation is being produced; and
(iv) Include an access control or locking
device that will prevent unauthorized use of the therapeutic radiation
machine;
(C) Windows,
mirrors, closed circuit television, or an equivalent viewing system must be
provided to permit continuous observation of the patient following positioning
and during irradiation and must be located so the operator may observe the
patient from the treatment control panel. The therapeutic radiation machine
must not be used for patient irradiation unless at least one viewing system is
operational;
(D) Provision must be
made for continuous two way aural communication between the patient and the
operator at the control panel. The therapeutic radiation machine must not be
used for irradiation of patients unless continuous two way aural communication
is possible;
(E) Treatment room
entrances must be provided with warning lights in a readily observable position
near the outside of all access doors that indicate when the useful beam is "ON"
and when it is "OFF";
(F)
Interlocks must be provided so all access controls are activated before
treatment can be initiated or continued. If the radiation beam is interrupted
by any access control, it must not be possible to restore the machine to
operation without resetting the access control and reinitiating irradiation by
manual action at the control panel;
(G) If the shielding material in any
protective barrier requires the presence of a beam interceptor to ensure
compliance with 180 NAC 4-013.01 and 4-013.02, interlocks must be provided to
prevent the production of radiation, unless the beam interceptor is in place,
whenever the useful beam is directed at the designated barrier or
barriers;
(H) At least one
emergency power cutoff switch must be located in the radiation therapy room and
must terminate all equipment electrical power including radiation and
mechanical motion. This switch is in addition to the termination switch
required by 180 NAC 20-007.11. All emergency power cutoff switches must include
a manual reset so that the therapeutic radiation machine cannot be restarted
from the unit's control console without resetting the emergency cutoff
switch;
(I) All safety interlocks
must be designed so that any defect or component failure in the safety
interlock system prevents or terminates operation of the therapeutic radiation
machine; and
(J) Surveys for
residual activity must be conducted on all therapeutic radiation machines
capable of generating photon and electron energies above 10 megavolts (MV)
prior to machining, removing, or working on therapeutic radiation machine
components which may have become activated due to photo-neutron
production.
007.18
RADIOLOGICAL MEDICAL PHYSICIST SUPPORT. The
requirements for radiological medical physicist support are as follows.
007.18(A)
RADIOLOGICAL MEDICAL
PHYSICIST RESPONSIBILITIES. The services of a radiological medical
physicist is required for registrants using therapeutic radiation machines with
energies of 500 kilovolts (kV) and above. The radiological medical physicist
must be responsible for:
(i) Full calibration
or calibrations required by 180 NAC 20-007.20 and protection surveys required
by 180 NAC 20-004.01;
(ii)
Supervision and review of dosimetry;
(iii) Beam data acquisition and transfer for
computerized dosimetry, and supervision of its use;
(iv) Quality assurance, including quality
assurance check review required by 180 NAC 20-007.21(E);
(v) Consultation with the user in treatment
planning, as needed; and
(vi)
Performing calculations and assessments regarding misadministrations.
007.18(B)
RADIOLOGICAL MEDICAL PHYSICIST CONTACT INFORMATION. If
the radiological medical physicist is not a full time employee of the
registrant, the operating procedures required by 180 NAC
20-007.19(F) must
also specifically address how the radiological medical physicist is to be
contacted for problems or emergencies, as well as the specific actions, if any,
to be taken until the radiological medical physicist can be
contacted.
007.19
OPERATING PROCEDURES. The requirements for operating
procedures are as follows.
007.19(A)
INDIVIDUALS OTHER THAN THE PATIENT. No individual,
other than the patient, must be in the treatment room during treatment or
during any irradiation for testing or calibration purposes.
007.19(B)
PROHIBITION. Therapeutic radiation machines must not
be made available for medical use unless the requirements of 180 NAC 20-004.01,
20-007.20 and
20-007.21 have been met.
007.19(C)
SECURED IF
UNATTENDED. Therapeutic radiation machines, when not in operation,
must be secured to prevent unauthorized use.
007.19(D)
RADIATION FIELD
POSITION AND SHAPE INDICATION. When adjustable beam limiting
devices are used, the position and shape of the radiation field must be
indicated by a light field.
007.19(E)
MECHANICAL SUPPORT OR
RESTRAINING DEVICES. If a patient must be held in position during
treatment, mechanical supporting or restraining devices must be used;
and
007.19(F)
OPERATING
AND EMERGENCY PROCEDURES. A copy of the current operating and
emergency procedures must be maintained at the therapeutic radiation machine
control console.
007.20
ACCEPTANCE TESTING, COMMISSIONING, AND FULL CALIBRATION
MEASUREMENTS. The requirements for acceptance testing,
commissioning, and full calibration measurements are as follows.
007.20(A)
PERFORMANCE. Acceptance testing, commissioning, and
full calibration of a therapeutic radiation machine subject to 180 NAC 20-007
must be performed by, or under the supervision of, a radiological medical
physicist.
007.20(B)
INSTALLATION OR REINSTALLATION. Acceptance testing and
commissioning must be performed according to "AAPM Code of Practice for
Radiotherapy Accelerators: Report of AAPM Radiation Therapy Task Group 45" and
manufacturers specifications, and must be conducted before the first medical
use following installation or reinstallation of the therapeutic radiation
machine. This document is available for viewing at the Department of Health and
Human Services, Division of Public Health, Office of Radiological Health, 301
Centennial Mall South, 3rd Floor, Lincoln, Nebraska 68509.
007.20(C)
FULL CALIBRATION
MEASUREMENTS. Full calibration must include measurement of all
applicable parameters required by Table II of "Comprehensive QA for Radiation
Oncology: Report of AAPM Radiation Therapy: AAPM Report No. 46," prepared by
Committee Task Group 40 and must be performed according to "AAPM Code of
Practice for Radiotherapy Accelerators: AAPM Report No. 47" prepared by
Radiation Therapy Task Group 45. These documents are available for viewing at
the Department of Health and Human Services, Division of Public Health, Office
of Radiological Health, 301 Centennial Mall South, 3rd Floor, Lincoln, Nebraska
68509. Although it may not be necessary to complete all elements of a full
calibration at the same time, all applicable parameters, for all energies, must
be completed at intervals not exceeding twelve calendar months, unless a more
frequent interval is required in Table II.
007.20(D)
FREQUENCY.
The radiological medical physicist must perform all elements of a full
calibration necessary to determine that all parameters are within acceptable
limits:
(i) Whenever quality assurance check
measurements indicate that the radiation output differs by more than 5 percent
from the value obtained at the last full calibration and the difference cannot
be reconciled. Therapeutic radiation machines with multi energy or multi-mode
capabilities must require measurements for only those modes or energies that
are not within their acceptable range; and
(ii) Following any component replacement,
major repair, or modification of components that could significantly affect the
characteristics of the radiation beam. If the repair, replacement or
modification does not affect all modes or energies, measurements must be
performed on the effected mode or energy that is in most frequent clinical use
at the facility. The remaining energies or modes may be validated with quality
assurance check procedures against the criteria in 180 NAC
20-007.20(D)(i).
007.20(E)
CALIBRATION
RECORDS. The registrant must maintain a record of each calibration
in an auditable form for the duration of the registration. The record must
include:
(i) The date of the calibration;
(ii) The manufacturer's name,
model number, and serial number for the therapeutic radiation machine;
(iii) The model numbers and serial
numbers of the instruments used to calibrate the therapeutic radiation machine;
and
(iv) The signature of the
radiological medical physicist responsible for performing the calibration.
007.21
PERIODIC QUALITY ASSURANCE CHECKS. The requirements
for periodic quality assurance checks are as follows.
007.21(A)
PERIODIC QUALITY
ASSURANCE CHECKS INTERVALS. Periodic quality assurance checks must
be performed on all therapeutic radiation machines subject to 180 NAC 20-007 at
intervals not to exceed those specified in "Comprehensive QA for Radiation
Oncology: AAPM Report No. 46," prepared by AAPM Radiation Therapy Committee
Task Group 40; or other procedure.
007.21(B)
QUALITY ASSURANCE CHECK
CONTENTS. To satisfy the requirement of 180 NAC
20-007.21(A),
quality assurance checks must include a determination of central axis radiation
output and a representative sampling of periodic quality assurance checks
contained in "Comprehensive QA for Radiation Oncology: AAPM Report No. 46"
prepared by Radiation Therapy Committee Task Group 40. The registrant may use
an alternative protocol that has been submitted to the Department for approval.
Representative sampling must include all applicable referenced periodic quality
assurance checks in an interval not to exceed 12 consecutive calendar
months.
007.21(C)
DOSIMETRY SYSTEM. The registrant must use a dosimetry
system that has been intercompared within the previous 12 months with the
dosimetry systems described in 180 NAC
20-004.03(A) to
make the periodic quality assurance checks required in 180 NAC
20-007.21(B).
007.21(D)
WRITTEN
PROCEDURES. The registrant must perform periodic quality assurance
checks required by 180 NAC
20-007.21(A)
according to written procedures established by the radiological medical
physicist;
007.21(E)
REVIEW OF RESULTS. The registrant must review the
results of each periodic radiation output check according to the following
procedures:
(i) The user and radiological
medical physicist must be immediately notified if any parameter is not within
its acceptable tolerance. The therapeutic radiation machine must not be made
available for subsequent medical use until the radiological medical physicist
has determined that all parameters are within their acceptable
tolerances;
(ii) If all quality
assurance check parameters appear to be within their acceptable range, the
quality assurance check must be reviewed and signed by either the user or
radiological medical physicist within 3 treatment days; and
(iii) The radiological medical physicist must
review and sign the results of each radiation output quality assurance check at
intervals not to exceed one month.
007.21(F)
QUALITY ASSURANCE CHECK
FREQUENCY. Therapeutic radiation machines subject to 180 NAC
20-007 must have applicable safety quality assurance checks listed in
"Comprehensive QA for Radiation Oncology: AAPM Report No. 46" prepared by AAPM
Radiation Therapy Committee Task Group 40. The registrant may use an
alternative protocol that has been submitted to the Department for approval.
The checks must be performed at intervals not to exceed 1 week.
007.21(G)
PROPER
OPERATION.To satisfy the requirement of 180 NAC
20-007.21(F),
safety quality assurance checks must ensure proper operation of:
(i) Electrical interlocks at each external
beam radiation therapy room entrance;
(ii) Proper operation of the "BEAM-ON",
interrupt and termination switches;
(iii) Beam condition indicator lights on the
access doors, control console, and in theradiation therapy room;
(iv) Viewing systems; and
(v) Electrically operated treatment room door
or doors from inside and outside the treatment room.
007.21(H) PROMPTLY REPAIR. The registrant
must promptly repair any system identified in 180 NAC
20-007.21(G) that
is not operating properly.
007.21(I)
QUALITY ASSURANCE CHECK
RECORDS. The registrant must maintain a record of each quality
assurance check required by 180 NAC
20-007.021(A) and
(G) for three years. The record must include:
(i) The date of the quality assurance
check;
(ii) The manufacturer's
name, model number, and serial number of the therapeutic radiation
machine;
(iii) The manufacturer's
name, model number, and serial number for the instrument or instruments used to
measure the radiation output of the therapeutic radiation machine;
and
(iv) The signature of the
individual who performed the periodic quality assurance check.
007.21(J)
QUALITY
ASSURANCE CHECKS FOR INTENSITY MODULATED RADIATION THERAPY (IMRT)
SYSTEMS. Quality assurance checks for intensity modulated
radiation therapy (IMRT) systems must:
(i)
Include commissioning and testing of the treatment planning and delivery
systems, routine quality assurance of the delivery system, and patient-specific
validation of treatment plans;
(ii)
Be performed according to "Guidance document on delivery, treatment planning,
and clinical implementation of IMRT: Report of the IMRT subcommittee of the
AAPM radiation therapy committee: AAPM Report No. 82". This document is
available for viewing at the Department of Health and Human Services, Division
of Public Health, Office of Radiological Health, 301 Centennial Mall South, 3rd
Floor, Lincoln, Nebraska 68509; and
(iii) Be performed according to the
manufacturer's specifications.