Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 180 - CONTROL OF RADIATION
Chapter 19 - LICENSES AND RADIATION SAFETY REQUIREMENTS FOR IRRADIATORS
Section 180-19-012 - CONTROL OF SOURCE MOVEMENT

Current through March 20, 2024

012.01 KEY. The mechanism that moves the sources of a panoramic irradiator must require a key to actuate. Actuation of the mechanism must cause an audible signal to indicate the sources are leaving the shielded position. Only one key may be in use at any time, and only operators or facility management may possess it. The key must be attached to a portable radiation survey meter by a chain or cable. The lock for source control must be designed so the key may not be removed if the sources are in an unshielded position. The door to the radiation room must require the same key.

012.02 POSITION DETECTOR. The console of a panoramic irradiator must have a source position indicator that indicates when the sources are in the fully shielded position, in transit, and exposed.

012.03 CONTROL CONSOLE. The control console of a panoramic irradiator must have a control that promptly returns the sources to the shielded position.

012.04 CONTROL MARKING. Each control for a panoramic irradiator must be clearly marked as to its function.

Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.