Nebraska Administrative Code
Topic - HEALTH AND HUMAN SERVICES SYSTEM
Title 179 - PUBLIC WATER SYSTEMS
Chapter 26 - REVISED TOTAL COLIFORM RULE
Section 179-26-009 - REPEAT MONITORING AND E. COLI REQUIREMENTS
Universal Citation: 179 NE Admin Rules and Regs ch 26 ยง 009
Current through September 17, 2024
009.01 Repeat Monitoring
26-009.01A If a sample taken under
26-005 through
26-008 is total coliform-positive,
the system must collect a set of repeat samples within 24 hours of being
notified of the positive result. The system must collect no fewer than three
repeat samples for each total coliform-positive sample found. The Department
may extend the 24-hour limit on a case-by-case basis if the system has a
logistical problem in collecting the repeat samples within 24 hours that is
beyond its control. Alternatively, the Department may implement criteria for
the system to use in lieu of case-by-case extensions. In the case of an
extension, the Department must specify how much time the system has to collect
the repeat samples. The Department cannot waive the requirement for a system to
collect repeat samples in
26-009.01A
through
26-009.01C.
26-009.01B The system must collect all repeat
samples on the same day, except that the Department may allow a system with a
single service connection to collect the required set of repeat samples over a
three-day period or to collect a larger volume repeat sample(s) in one or more
sample containers of any size, as long as the total volume collected is at
least 300 ml.
26-009.01C The system
must collect an additional set of repeat samples in the manner specified in
26-009.01A
through
26-009.01C if one
or more repeat samples in the current set of repeat samples is total
coliform-positive. The system must collect the additional set of repeat samples
within 24 hours of being notified of the positive result, unless the Department
extends the limit as provided in
26-009.01A. The
system must continue to collect additional sets of repeat samples until either
total coliforms are not detected in one complete set of repeat samples or the
system determines that a coliform treatment technique trigger specified in
26-010.01 has been exceeded as a result of a repeat sample being total
coliform-positive and notifies the Department. If a trigger identified in
26-010 is exceeded as a result of a
routine sample being total coliform-positive, systems are required to conduct
only one round of repeat monitoring for each total coliform-positive routine
sample.
26-009.01D After a system
collects a routine sample and before it learns the results of the analysis of
that sample, if it collects another routine sample(s) from within five adjacent
service connections of the initial sample, and the initial sample, after
analysis, is found to contain total coliforms, then the system may count the
subsequent sample(s) as a repeat sample instead of as a routine
sample.
26-009.01E Results of all
routine and repeat samples taken under
26-005 through 26-009 not invalidated
by the Department must be used to determine whether a coliform treatment
technique trigger specified in
26-010 has been exceeded.
009.02 Escherichia coli (E. coli) Testing
26-009.02A If any routine or repeat
sample is total coliform-positive, the system must analyze that total
coliform-positive culture medium to determine if E. coli are
present. If E. coli are present, the system must notify the
Department by the end of the day when the system is notified of the test
result, unless the system is notified of the result after the Department office
is closed and the Department does not have either an after-hours phone line or
an alternative notification procedure, in which case the system must notify the
Department before the end of the next business day.
26-009.02B The Department has the discretion
to allow a system, on a case-by-case basis, to forgo E. coli
testing on a total coliform-positive sample if that system assumes
that the total coliform-positive sample is E. coli-positive.
Accordingly, the system must notify the Department as specified in
26-009.02A and
the provisions of 179 NAC
2-002.04C3
apply.
Disclaimer: These regulations may not be the most recent version. Nebraska may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.