Administrative Rules of Montana
Department 37 - PUBLIC HEALTH AND HUMAN SERVICES
Chapter 37.14 - RADIATION CONTROL
Subchapter 37.14.13 - Radiation Safety Requirements for Analytical X-Ray Equipment
Rule 37.14.1306 - AREA REQUIREMENTS
Current through Register Vol. 18, September 20, 2024
(1) The local components of an analytical x-ray system shall be located and arranged and shall include sufficient shielding or access control such that no radiation levels exist in any area surrounding the local component group which could result in a dose to an individual present therein in excess of the dose limits given in ARM 37.14.709. For systems utilizing x-ray tubes, these levels shall be met at any specified tube rating.
(2) Radiation surveys, as required by ARM 37.14.720, shall be performed of all analytical x-ray systems sufficient to show compliance with (1) of this rule unless a registrant can demonstrate to the satisfaction of the department compliance in some other manner. These surveys shall be performed:
(3) Each area or room containing analytical x-ray equipment shall be conspicuously posted with a sign or signs bearing the radiation symbol and the words "CAUTION--X-RAY EQUIPMENT", or words having a similar intent.
Sec. 75-3-201 and 75-3-202, MCA; IMP, Sec. 75-3-201 and 75-3-202, MCA;