Rule 37.14.1021 - SPECIAL REQUIREMENTS FOR CHEST PHOTOFLUOROGRAPHIC INSTALLATIONS
Universal Citation: MT Admin Rules 37.14.1021
Current through Register Vol. 18, September 20, 2024
(1) The equipment used in chest photofluorographic installations shall be as follows:
(a) All requirements of ARM
37.14.1012(1)
shall be satisfied.
(b) A
collimator shall restrict the useful beam to the area of the photofluorographic
screen.
(2) All requirements of ARM 37.14.1003(5) and 37.14.1012(2) concerning structural shielding shall be satisfied.
(3) The operating procedures applicable to chest photofluorographic installations shall be as follows:
(a) All requirements of ARM
37.14.1012(3)
shall be satisfied.
(b) All
individuals except the patient being examined shall be in shielded positions
during exposures.
(c) Personnel
monitoring shall be required for all individuals associated with the operation
of the equipment.
Sec. 75-3-201 and 75-3-202, MCA; IMP, Sec. 75-3-201, 75-3-202 and 75-3-204, MCA;
Disclaimer: These regulations may not be the most recent version. Montana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.