Rule 37.14.1020 - SPECIAL REQUIREMENTS FOR MOBILE DIAGNOSTIC RADIOGRAPHIC EQUIPMENT
Universal Citation: MT Admin Rules 37.14.1020
Current through Register Vol. 18, September 20, 2024
(1) The equipment used in mobile diagnostic radiography shall be as follows:
(a) All requirements of ARM
37.14.1012(1)
shall be satisfied except (g).
(b)
A type of dead-man switch for an exposure switch shall be provided, together
with electrical cord of sufficient length so that the operator can stand at
least 6 feet from the useful beam.
(2) When a mobile unit is used routinely in 1 location, it shall be considered a fixed installation subject to the shielding requirements specified in ARM 37.14.1003(5) and 37.14.1012(2).
(3) The operating procedures applicable to mobile diagnostic radiographic equipment are as follows:
(a) All provisions of ARM
37.14.1012(3)
apply.
(b) The target-to-skin
distance shall not be less than 15 inches.
(c) Personnel monitoring shall be required
for all persons associated with the operation of mobile x-ray
equipment.
(d) Mobile x-ray
equipment shall not be used for photofluoroscopy unless it meets the
requirements for mobile fluoroscopes in ARM
37.14.1006.
(e) The operator shall wear a protective
apron or stand behind a suitable shield.
Sec. 75-3-201 and 75-3-202, MCA; IMP, Sec. 75-3-201, 75-3-202 and 75-3-204, MCA;
Disclaimer: These regulations may not be the most recent version. Montana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.