Administrative Rules of Montana
Department 37 - PUBLIC HEALTH AND HUMAN SERVICES
Chapter 37.14 - RADIATION CONTROL
Subchapter 37.14.10 - X-Rays in the Healing Arts
Rule 37.14.1012 - RADIOGRAPHIC INSTALLATIONS OTHER THAN DENTAL AND VETERINARY MEDICINE
Universal Citation: MT Admin Rules 37.14.1012
Current through Register Vol. 18, September 20, 2024
(1) The equipment used in radiographic installations in other than dental and veterinary medicine shall be as follows:
(a) The tube housing
will be of diagnostic type.
(b)
Suitable devices (diaphragms, cones, adjustable collimators), capable of
restricting the useful beam to the area of clinical interest shall be provided
to define the beam and shall provide the same degree of attenuation as that
required of the tube housing. Such devices shall be calibrated in terms of the
size of the projected useful beam at specified source-film distances. For chest
photofluorographic equipment, the collimator shall restrict the beam to
dimensions no greater than those of the fluorographic screen and shall confine
the beam to the screen.
(c)
Radiographic equipment, including multipurpose machines sold in Montana after
July 1, 1969, shall be equipped with adjustable rectangular collimators
containing light localizers that define the entire field. Equipment installed
prior to March 28, 1980, utilizing circular collimators which have met or will
meet established standards will be acceptable. The field size indication on
adjustable collimators shall be accurate with a tolerance not exceeding 1 inch
for a source-film distance of 72 inches, the light field shall align with the
x-ray field with the same degree of accuracy.
(d) The size of the x-ray beam projected by
fixed aperture cones and collimators, except those used for stereo-radiography,
should not exceed the minimum dimensions of the x-ray film by more than 2
inches (1 inch border) for a source-film distance of 72 inches or 1 inch (1/2
inch border) for a source-film distance of 36 inches.
(e) Except when contraindicated for a
particular radiographic purpose, the aluminum equivalent of the total
filtration in the primary beam shall not be less than 0.5 millimeters aluminum
for machines capable of operating not over 50 kVp, 1.5 millimeters for those
capable of operating between 50-70 kVp and 2.5 millimeters for those capable of
operating above 70 kVp.
(f) A
device shall be provided which terminates the exposure at a preset time
interval or exposure. The operator shall be able to terminate the exposure at
any time.
(g) A type of dead-man
switch for an exposure switch, except for those used in conjunction with
"spot-film", shall be required and so arranged that it cannot be operated
outside a shielded area.
(h) The
control panel shall include a device such as a milliammeter or equivalent to
give positive indication of the production of x-rays whenever the x-ray tube is
energized.
(i) The control panel
shall include devices, (labeled control settings and/or meters) indicating the
physical factors (such as kVp, mA, exposure time, or whether timing is
automatic) used for the exposure.
(j) Machines equipped with beryllium window
x-ray tubes shall contain keyed filter interlock switches in the tube housing
and suitable indication on the control panel of the added filter in the useful
beam if the total filtration permanently in the useful beam is less than 0.5 mm
aluminum equivalent. The total filtration permanently in the useful beam shall
be clearly indicated on the tube housing.
(k) Beryllium window x-ray tubes shall not be
used on multipurpose radiographic equipment.
(l) Gonadal shielding shall be used for the
patient when appropriate.
(m) When
a patient must be held in position for radiography, mechanical supporting or
restraining devices shall be used whenever possible. If the patient must be
held by an individual, that individual shall be protected with appropriate
shielding devices such as protective gloves and apron and he shall be so
positioned that no part of his body will be struck by the useful beam and that
his body is as far as possible from the edge of the useful beam.
(n) Only persons whose presence is necessary
shall be in the radiographic room during exposure. All such persons shall be
protected.
(2) The structural shielding of radiographic equipment in other than dental or veterinary medicine shall be as follows:
(a)
All wall, floor, and ceiling areas exposed to the useful beam shall have
primary barriers. Primary barriers in walls shall extend to a minimum height of
84 inches above the floor.
(b)
Secondary barriers shall be provided in all wall, floor, and ceiling areas not
having primary barriers or where the primary barrier requirements are lower
than the secondary barrier requirements.
(c) The operator's station at the control
shall be behind a protective barrier, either in a separate room, in a protected
booth, or behind a shield which will intercept the useful beam and any
radiation which has been scattered only once.
(d) A window of lead-equivalent glass equal
to that required by the adjacent barrier or a mirror system shall be provided
large enough and so placed that the operator can see the patient without having
to leave the protected area during exposure.
(3) The operating procedures applicable to radiographic installations other than in dental or veterinary medicine shall be as follows:
(a) Personnel monitoring shall be
performed in controlled areas for each occupationally exposed individual beyond
the required 13-week period, for whom there is a reasonable possibility of
receiving a dose exceeding one-fourth the applicable maximum permissible
dose.
(b) No individual
occupationally exposed to radiation shall be permitted to hold patients during
exposures except during emergencies, nor shall any individual be regularly used
for this service.
(c) Only
individuals required for the radiographic procedure shall be in the
radiographic room during exposure; and, except for the patient, no unprotected
parts of their bodies shall be in the useful beam.
(d) The useful beam shall be restricted to
the area of clinical interest.
Sec. 75-3-201 and 75-3-202, MCA; IMP, Sec. 75-3-201, 75-3-202 and 75-3-204, MCA;
Disclaimer: These regulations may not be the most recent version. Montana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google
Privacy Policy and
Terms of Service apply.