Rule 37.14.1010 - DENTAL RADIOGRAPHIC INSTALLATIONS
Universal Citation: MT Admin Rules 37.14.1010
Current through Register Vol. 18, September 20, 2024
(1) The equipment used in dental radiographic installations shall be as follows:
(a) The
tube housing shall be of diagnostic type.
(b) Diaphragms or cones shall be used for
collimating the useful beam and provide the same degree of protection as the
housing. The diameter of the useful beam at the cone tip shall not be more than
2-3/4 inches for intra-oral radiography.
(c) A cone or spacer frame shall provide a
target-to-skin distance of not less than 7 inches with apparatus operating
above 50 kVp or 4 inches with apparatus operating at 50 kVp or below.
(d) For equipment capable of operating up to
no more than 70 kVp, the total filtration permanently in the useful beam shall
be equivalent to at least 1.5 mm of aluminum. This requirement may be assumed
to have been met if the half-value layer is not less than 1.5 mm aluminum at
maximum possible operating voltages.
(e) For equipment capable of operating above
70 kVp, the total filtration permanently in the useful beam shall be equivalent
to at least 2.5 mm of aluminum. This requirement may be assumed to have been
met if the half-value layer is not less than 2.5 mm aluminum at the maximum
possible operating voltage.
(f)
Means shall be provided to terminate the exposure at a preset time interval,
preset product of current and time, a preset number of pulses, or a preset
radiation exposure to the image receptor. In addition:
(i) It shall not be possible to make an
exposure when the timer is set to a "zero" or "off" position if either position
is provided.
(ii) With a timer
setting of 0.5 seconds or less, the average exposure period (T) shall be
greater than or equal to 5 times the maximum exposure period
(Tmax) minus the minimum exposure period
(Tmin) when 4 timer tests are performed, i.e.,
T > 5(Tmax - Tmin).
(g) An x-ray control shall be incorporated
into each x-ray system such that an exposure can be terminated by the operator
at any time, except for exposure of one-half second or less.
(h) Each installation shall be provided with
a protective barrier for the operator or shall be so arranged that the operator
can stand at least 12 feet from the useful beam. Stationary x-ray systems
installed after July 1, 1980, shall be required to have the x-ray control
permanently mounted in a protected area, e.g., corridor outside the room, so
that the operator is required to remain in that protected area during the
entire exposure.
(2) The structural shielding requirements for dental radiographic installations shall be as follows:
(a) Dental rooms containing
x-ray machines shall be provided with primary protective barriers at all areas
struck by the useful beam. Consideration shall be given to the attenuation
provided by the patient and the building materials in partitions.
(b) When dental x-ray units are installed in
adjacent rooms or areas, protective barriers shall be provided between the
rooms or areas. Consideration of the effects of the separating wall(s) shall be
given in the evaluation of this protection.
(3) The operating procedures applicable to dental radiographic installations shall be as follows:
(a) Neither the dentist nor his assistant
shall be permitted to hold patients or films during exposure, nor shall any
individual be regularly used for this service.
(b) During each exposure, the operator shall
stand at least 12 feet from the patient or behind a protective
barrier.
(c) Only the patient shall
be in the useful beam.
(d) Neither
the tube housing nor the pointer cone shall be hand-held during
exposure.
(e) Fluoroscopy shall not
be used in dental examinations.
(f)
Personnel monitoring shall be performed in controlled areas for each
occupationally exposed individual beyond the required 13-week period, for whom
there is a reasonable possibility of receiving a dose exceeding one-fourth the
applicable maximum permissible dose.
Sec. 75-3-201 and 75-3-202, MCA; IMP, Sec. 75-3-201 and 75-3-202, MCA;
Disclaimer: These regulations may not be the most recent version. Montana may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.