Section 20 CSR 2267-5.030 - Cleaning and Sterilization
Universal Citation: 20 MO Code of State Regs 2267-5.030
Current through Register Vol. 49, No. 18, September 16, 2024
PURPOSE: This rule outlines the proper cleaning and sterilization of equipment used by a tattooist, body piercer or brander.
(1) Cleaning.
(A) Reusable equipment used in a tattooing,
body piercing and branding procedure shall be cleaned immediately following
each use to remove blood and/or body tissue residue prior to
sterilization.
(B) Reusable bars,
tubes, branding irons, other branding instruments and body piercing equipment
shall be placed in a covered container until they are sterilized.
(C) All containers holding contaminated
tubes, branding irons, other branding equipment, reusable body piercing
equipment, and container lids shall be cleaned and disinfected with an approved
disinfectant as defined in
20 CSR
2267-1.010.
(D) Gloved personnel shall clean bars, tubes,
branding irons, other branding equipment, and reusable body piercing equipment
prior to sterilization as follows:
1.
Manually preclean the items with care taken to ensure removal of residue;
thoroughly rinse the items with warm water and then drain the water; clean the
items by soaking them in a protein-dissolving detergent-enzyme cleaner used
according to the manufacturer's instructions; and clean the items further in an
ultrasonic cleaning device that operates at forty to sixty (40-60) hertz and is
used according to the manufacturer's instructions; and
2. Rinsing and drying the items.
(E) Prior to autoclaving, all
tubes shall be packaged either individually or in quantities appropriate for
individual procedures. Packages shall be identifiable and dated.
(2) Sterilization.
(A) Equipment requiring sterilization shall
be pressure-sterilized at the establishment in an autoclave and in accordance
with manufacturer's instructions. Practitioners shall have procedures in place
to ensure autoclaves have been properly disinfected and spore-tested as
required in subsection (2)(C) of this rule.
(B) Each batch of sterilized equipment shall
be monitored for sterilization by use of heat-sensitive indicators capable of
indicating approximate time and temperature achieved.
(C) Autoclaves shall be spore-tested at least
weekly. Spore kill test effectiveness shall be conducted by an independent
laboratory. If a positive spore test is received, the practitioner shall
immediately cease using the autoclave device and notify the office within
forty-eight (48) hours.
(D)
Sterilized equipment shall be wrapped and stored in a manner that ensures it
will remain sterile until used.
(E)
Each tattoo, body piercing and branding establishment shall maintain
sterilization records including spore tests for at least two (2) years from the
date of the last entry, which shall include the following information:
1. Date of sterilization;
2. Name of person operating the equipment;
and
3. Result of heat-sensitive
indicator.
(F)
Sterilized equipment shall be resterilized if the package is opened, damaged or
becomes wet.
(G) All methods of
sterilization other than steam autoclaving are prohibited.
*Original authority: 324.522, RSMo 1998, amended 1999, 2001, 2007.
Disclaimer: These regulations may not be the most recent version. Missouri may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.