Current through Register Vol. 49, No. 18, September 16, 2024
(1) All variable
parameters of the equipment shall be calibrated as follows:
(A) When the equipment is first
installed;
(B) After any major
changes or replacement of parts;
(C) At least annually while in use;
and
(D) When quality assurance
tests indicate that calibration is needed.
(2) The following parameters of the system
shall be monitored as follows:
(A) Processor
performance shall be monitored on a daily basis to include solution
temperatures, sensitometric-densitometric evaluation through the use of control
charts which shall include speed step or mid-density step with a control limit
of plus or minus fifteen-hundredths optical density (+-0.15 OD), contrast index
or density difference with a control limit of plus or minus fifteen-hundredths
optical density (+-0.15 OD) and base plus fog with a control limit of plus
three-hundredths optical density (+.03 OD) of the operating level;
(B) Processor chemical replenishment rates
shall be monitored at least quarterly;
(C) Half-value layer shall be determined at
least annually with the compression device located in the primary beam, halfway
between the image receptor assembly and the X-ray tube;
(D) Milliampere-seconds (mAs) linearity shall
be determined at least annually based on the average ratios of exposure to the
indicated milliampere-seconds product (mR/mAs) obtained at any two (2)
consecutive tube current settings, which shall not differ by more than
ten-hundredths (0.10) times their sum as measured with a minimum of four (4)
consecutive exposures or a maximum of ten (10) consecutive exposures at
commonly used settings;
(E) Manual
and automatic exposure control (AEC) reproducibility shall be determined at
least annually based on the coefficient of variation of exposures, which shall
not be greater than plus or minus five percent (+-5%) as measured with a
minimum of four (4) consecutive exposures or a maximum of ten (10) consecutive
exposures at a commonly used setting;
(F) Kilovoltage potential (kVp) accuracy
shall be determined at least annually. The actual kVp shall be maintained
within plus or minus five percent (+-5%) of the indicated kVp;
(G) kVp reproducibility shall be determined
at least annually based on the coefficient of variation of kVp values, which
shall not be greater than plus or minus five percent (+-5%) as measured with a
minimum of four (4) consecutive exposures or a maximum of ten (10) consecutive
exposures at commonly used kVp settings;
(H) AEC thickness response shall be
determined at least annually based on film densi-t y, which shall be maintained
within plus or minus three-tenths optical density (+-0.3 OD) of the average OD
over the kVp range used for phantom thicknesses of two, four and six
centimeters (2, 4 and 6 cm). If the OD cannot be maintained to within plus or
minus three-tenths (00.3) of the average of each clinically used setting, a
technique chart shall be developed that alters kVp and density control settings
as a function of breast thicknesses and densities to produce optical densities
within this range under phototimed conditions;
(I) Compression device response shall be
monitored at least annually;
(J)
Adequacy of film storage shall be monitored at least annually, both before use
and after exposure, if processing does not occur immediately;
(K) Availability and use of technique charts
that shall include an indication of the kV-tar-get-filter combination to be
used with each image receptor shall be monitored at least annually;
(L) The use of kV-target-filter combination
appropriate to the image receptor shall be monitored daily before patient
irradiation;
(M) Darkroom
integrity, for light-tight conditions and use of proper safelight, shall be
monitored at least semiannually and whenever bulbs or filters are changed or
when fog is suspected; darkroom fog levels shall have a variance of not greater
than five-hundredths (0.05) density units between the fogged and unfogged film
based on a two (2)-minute exposure to the darkroom safelight;
(N) Image quality, using one (1) of the
following mammography phantoms; the RMI 152, the RMI 156, the CIRS, the Nuclear
Associates phantom or other equivalent phantom acceptable to the Department of
Health, which simulates the composition of the breast and provides indicators
of disease conditions allowing objective analysis of clinical image quality,
shall be monitored at least monthly and each time the mammographic equipment is
moved or altered in any major way including the replacement of parts; a phantom
image shall be made and evaluated for mobile units and vans at each location
prior to the performance of mammography;
(O) Average glandular dose shall be
determined at least annually based on measurements of estimated skin entrance
exposure; the average glandular dose shall be calculated from those used for
imaging a four and five-tenths centimeters (4.5 cm) compressed breast in the
cranial caudal view and the measurement point shall be four and five-tenths
centimeters (4.5 cm) from the surface of the grid or cassette as appropriate;
and
(P) Film-screen contact shall
be monitored at least semiannually and when new or repaired cassettes/screens
are placed in service, using a contact test tool specifically designed for
mammography; cassettes/ screens shall be cleaned at least weekly; for
xeroradiography, the photoreceptor plates shall be maintained in accordance
with the manufacturer's specifications.
(3) Evaluation of monitoring results-
(A) Standards of image quality giving
acceptable ranges of value for each of the parameters tested shall be
established to aid in the evaluation of monitoring results. The standards of
image quality related to dose shall include a requirement that the average
glandular dose for one (1) cranial caudal view of a four and five-tenths
centimeters (4.5 cm) compressed breast fifty percent (50%) glandular/fifty
percent (50%) adipose composition shall not exceed one hundred (100), three
hundred (300) and four hundred (400) millirads (mrads) for film/screen units
without grids, film/screen units with grids, and xeroradiography units,
respectively; and
(B) The
monitoring results shall be compared routinely to the standards of image
quality established by the Department of Health. If the results fall outside
the acceptable range, the test shall be repeated. If the results continue to be
unacceptable, the source of the problem shall be identified and corrected
before further examinations are conducted.
(4) A program to analyze retakes shall be
established to detect and correct problems affecting image quality or exposure.
This analysis shall be performed at least quarterly or after two hunndred fifty
(250) patients, whichever comes first.
(5) All quality assurance test equipment
shall be in good operating order and calibrated according to the manufacturer's
specifications.
*Original authority
1992.