Section 19 CSR 20-10.140 - Special Requirements for Medical Radiographic Installations
Universal Citation: 19 MO Code of State Regs 20-10.140
Current through Register Vol. 49, No. 18, September 16, 2024
PURPOSE: This rule establishes special requirements for medical radiographic X-ray installations.
(1) A device shall be provided to terminate the exposure after a preset time or exposure.
(2) The exposure switch shall be so arranged that it cannot be conveniently operated outside a shielded area. Exposure switches for spot film devices used in conjunction with fluoroscopic tables are excepted from this requirement.
*Original authority: 192.420, RSMo 1963.
Disclaimer: These regulations may not be the most recent version. Missouri may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.