Mississippi Administrative Code
Title 15 - Mississippi Department of Health
Part 21 - Division of Radiological Health
Subpart 78 - Radiological Health
Chapter 1 - REGULATIONS FOR CONTROL OF RADIATION IN MISSISSIPPI
Subchapter 9 - Radiation Safety Requirements for Particle Accelerators
Rule 15-21-78-1.9.8 - Particle Accelerator Controls and Interlock Systems

Current through September 24, 2024

1. Instrumentation, readouts, and controls on the particle accelerator control console shall be clearly identified and easily discernible.

2. Each entrance into a target room or other high radiation area shall be provided with a safety interlock that shuts down the machine under conditions of barrier penetration.

3. Each safety interlock shall be on a circuit which shall allow it to operate independently of all other safety interlocks.

4. All safety interlocks shall be designed so that any defect or component failure in the safety interlock system prevents operation of the accelerator.

5. When a safety interlock system has been tripped, it shall only be possible to resume operation of the accelerator by manually resetting controls at the position where the safety interlock has been tripped and, lastly, at the main control console.

6. A scram button or other emergency power cutoff switch shall be located and easily identifiable in all high radiation areas. Such a cutoff switch shall include a manual reset so that the accelerator cannot be restarted from the accelerator control console without resetting the cutoff switch.

Miss. Code Ann. § 45-14-11

Disclaimer: These regulations may not be the most recent version. Mississippi may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.
This site is protected by reCAPTCHA and the Google Privacy Policy and Terms of Service apply.