Rule 15-21-78-1.9.10 - Operating Procedures
Current through September 24, 2024
1. Particle accelerators, when not in operation, shall be secured to prevent unauthorized use.
2. The safety interlock system shall not be used to turn off the accelerator beam except in an emergency.
3. All safety and warning devices, including interlocks, shall be checked for proper operation at intervals not to exceed three months. Results of such tests shall be maintained at the accelerator facility for inspection by the Agency.
4. Electrical circuit diagrams of the accelerator and the associated safety interlock systems shall be kept current and maintained for inspection by the Agency and shall be available to the operator at each accelerator facility.
5. If, for any reason, it is necessary to intentionally bypass a safety interlock or interlocks, such action shall be:
6. A copy of the current operating and the emergency procedures shall be maintained at the accelerator control panel.
Miss. Code Ann. § 45-14-11