Rule 15-21-78-1.8.3 - Equipment Requirements
Current through September 24, 2024
1. Safety Device. A device which prevents the entry of any portion of an individual's body into the primary x-ray beam path or which causes the beam to be shut off upon entry into its path shall be provided on all open- beam configurations. A registrant may apply to the Agency for an exemption from the requirements of a safety device. Such application shall include:
2 Warning Devices
3. Ports. Unused ports on radiation source housings shall be secured in the closed position in a manner which will prevent casual opening.
4. Labeling. All analytical x-ray equipment shall be labeled with a readily discernible sign or signs bearing the radiation symbol and the words:
5. Shutters. On open-beam configurations installed after January 12, 1980, each port on the radiation source housing shall be equipped with a shutter that cannot be opened unless a collimator or a coupling has been connected to the port.
6. Radiation Source Housing. Each radiation source housing shall be subject to the following requirements:
7. Generator Cabinet. Each x-ray generator shall be supplied with a protective cabinet which limits leakage radiation measured at a distance of 5 centimeters from its surface such that it is not capable of producing a dose in excess of 0.25 millirem (2.5 µSv) in one hour.