A. Reference standards that are used in the laboratory must be obtained, when available, from the National Institute of Standards and Technology (NIST), manufacturers that supply NIST standards or NIST traceable standards, or an international standard-setting organization.

B. A laboratory must retain records for all standards, reagents, and bacteriological media. The records must include:

(1) identification of the manufacturer or vendor;

(2) certificate of analysis or purity, if supplied;

(3) lot number;

(4) date of receipt or preparation;

(5) preparer's initials, if applicable;

(6) method of preparation, when prepared in the laboratory;

(7) recommended storage conditions; and

(8) expiration date after which the material must not be used unless its reliability is verified by the laboratory.

Statutory Authority: MS s 144.97; 144.98

Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.