Part 4732.1063 - WARNING DEVICES FOR INDUSTRIAL RADIOGRAPHY FACILITIES

Universal Citation: MN Rules 4732.1063

Current through Register Vol. 48, No. 39, March 25, 2024

Subpart 1. Open-beam configurations.

Open-beam configurations must have a readily discernible indication of:

A. x-ray tube "on-off" status located near the radiation-producing equipment housing, if the primary beam is controlled in an "on-off" manner; or

B. shutter "open-closed" status located near each port on the radiation-producing equipment housing, if the primary beam is controlled in "open-closed" manner.

Subp. 2. Warning light.

An easily visible warning light labeled with the words "X-RAY ON" or other visible warning indicator that clearly shows the equipment is producing ionizing radiation, must be located near a switch that energizes an x-ray tube and illuminated only when the tube is energized.

Subp. 3. Warning device labeling.

Warning devices must be labeled so that their purpose is easily identified.

Statutory Authority: MS s 144.12

Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.