Minnesota Administrative Rules
Agency 144 - Health Department
Chapter 4731 - RADIATION SAFETY
MEDICAL USE OF RADIOACTIVE MATERIAL
Part 4731.4520 - SPOT CHECK RECORDS; REMOTE AFTERLOADER UNITS
Universal Citation: MN Rules 4731.4520
Current through Register Vol. 49, No. 13, September 23, 2024
A. A licensee must retain a record of each spot check for remote afterloader units required under part 4731.4473 for three years. The record must include, as applicable:
(1) the date of the spot check;
(2) the unit manufacturer's name;
(3) the model number and serial number for
the remote afterloader unit;
(4)
the model number, serial number, and identity of the source;
(5) an assessment of timer
accuracy;
(6) notations indicating
the operability of each entrance door electrical interlock, radiation monitor,
source exposure indicator light, viewing and intercom system, and clock and
decayed source activity in the unit's computer;
(7) the name of the individual who performed
the periodic spot check; and
(8)
the signature of the authorized medical physicist who reviewed the record of
the spot check.
B. A licensee must retain a copy of the procedures required under part 4731.4473, subpart 2, until the licensee no longer possesses the remote afterloader unit.
Statutory Authority: MS s 144.1202; 144.1203
Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.