Minnesota Administrative Rules
Agency 144 - Health Department
Chapter 4731 - RADIATION SAFETY
MEDICAL USE OF RADIOACTIVE MATERIAL
Part 4731.4475 - MOBILE REMOTE AFTERLOADER UNITS; ADDITIONAL REQUIREMENTS
Current through Register Vol. 49, No. 13, September 23, 2024
Subpart 1. General requirements.
A licensee providing mobile remote afterloader service must:
Subp. 2. Check requirements.
In addition to the periodic spot checks required under part 4731.4473, a licensee authorized to use mobile afterloaders for medical use must perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
Subp. 3. Simulated treatment cycle.
In addition to the requirements for checks under subpart 2, a licensee must ensure proper overall operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
Subp. 4. Malfunctions.
If the results of the checks required under subpart 2 indicate the malfunction of any system, a licensee must lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
Subp. 5. Record retention.
A licensee must retain a record of each check required under subpart 2 according to part 4731.4522.
Statutory Authority: MS s 144.1202; 144.1203