Minnesota Administrative Rules
Agency 144 - Health Department
Chapter 4731 - RADIATION SAFETY
MEDICAL USE OF RADIOACTIVE MATERIAL
Part 4731.4470 - REMOTE AFTERLOADER UNITS; FULL CALIBRATION
Current through Register Vol. 49, No. 13, September 23, 2024
Subpart 1. Calibration required.
A licensee authorized to use a remote afterloader unit for medical use must perform full calibration measurements on each unit:
Subp. 2. Required determinations.
To satisfy subpart 1, full calibration measurements must include, as applicable, determination of:
Subp. 3. Required system.
A licensee must use the dosimetry system described in part 4731.4468, subpart 1, to measure the output.
Subp. 4. Required protocols.
A licensee must make full calibration measurements required under subpart 1 according to published protocols accepted by nationally recognized bodies.
Subp. 5. Autoradiograph required.
In addition to the requirements for full calibrations for low dose-rate remote afterloader units under subpart 2, a licensee must perform an autoradiograph of the source to verify inventory and source arrangement at intervals not exceeding one quarter.
Subp. 6. Measurements by manufacturer.
For low dose-rate remote afterloader units, a licensee may use measurements provided by the source manufacturer that are made according to subparts 1 to 5.
Subp. 7. Required corrections.
A licensee must mathematically correct the outputs determined in subpart 2, item A, for physical decay at intervals consistent with one percent physical decay.
Subp. 8. Authorized medical physicist.
Full calibration measurements required under subpart 1 and physical decay corrections required under subpart 7 must be performed by the authorized medical physicist.
Subp. 9. Record retention.
A licensee must retain a record of each calibration according to part 4731.4518.
Statutory Authority: MS s 144.1202; 144.1203