Minnesota Administrative Rules
Agency 144 - Health Department
Chapter 4731 - RADIATION SAFETY
MEDICAL USE OF RADIOACTIVE MATERIAL
Part 4731.4455 - BRACHYTHERAPY; CALIBRATION MEASUREMENTS

Universal Citation: MN Rules 4731.4455

Current through Register Vol. 49, No. 13, September 23, 2024

A. Before the first medical use of a brachytherapy source, a licensee must have:

(1) determined the source output or activity using a dosimetry system that meets the requirements of part 4731.4468, subpart 1;

(2) determined source positioning accuracy within applicators; and

(3) used published protocols currently accepted by nationally recognized bodies to meet the requirements of this item.

B. A licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine that are made according to item A.

C. A licensee must mathematically correct the outputs or activities determined under item A for physical decay at intervals consistent with one percent physical decay.

D. A licensee must retain a record of each calibration according to part 4731.4513.

Statutory Authority: MS s 144.1202; 144.1203

Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.