Minnesota Administrative Rules
Agency 144 - Health Department
Chapter 4731 - RADIATION SAFETY
MEDICAL USE OF RADIOACTIVE MATERIAL
Part 4731.4421 - CALIBRATION OF SURVEY INSTRUMENTS
Universal Citation: MN Rules 4731.4421
Current through Register Vol. 49, No. 13, September 23, 2024
A. A licensee must calibrate the survey instruments used to show compliance with parts 4731.2000 to 4731.2950 and 4731.4400 to 4731.4527 before first use, intervals not to exceed 12 months, and following a repair that affects the calibration. A licensee must:
(1) calibrate
all scales with readings up to 1,000 millirems (10 mSv) per hour with a
radiation source;
(2) calibrate two
separate readings on each scale or decade that will be used to show compliance;
and
(3) conspicuously note on the
instrument the date of calibration.
B. A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.
C. A licensee must retain a record of each survey instrument calibration according to part 4731.4502, subpart 2.
Statutory Authority: MS s 144.1202; 144.1203
Disclaimer: These regulations may not be the most recent version. Minnesota may have more current or accurate information. We make no warranties or guarantees about the accuracy, completeness, or adequacy of the information contained on this site or the information linked to on the state site. Please check official sources.